Braun 950787 manual Vtbd and Time, Rate and Time, To Change the Rate While Infusing

Page 10
to begin the infusion. Check for drops in the drip chamber.

Continuous Mode

chamber.

VTBD and TIME

1.Press VTBD.

2.Enter bag volume in ml.

3.Press OK. (To leave screen without saving changes, press EXIT.)

4.Press TIME.

5.Enter infusion time in hours and minutes.

NOTE: For a 2 hour infusion, enter 2, 0, 0. The pump will show: Time: 02h00m

6.Press OK. (To leave screen without saving changes, press EXIT.) The pump calculates the rate and asks: Accept Rate?

7.Press OK.

8. Press

RATE and TIME

1.Enter rate in ml/hr.

2.Press TIME.

3.Enter infusion time in hours and minutes.

4.Press OK. (To leave screen without saving changes, press EXIT.)

 

5. Press

to begin the infusion. Check for drops in the drip

 

 

chamber.

To Change the Rate

1.

Press

to stop the infusion.

2.

Press

. The display will reset to 000.0 ml/h.

3.

Enter a new rate.

4.

Press

to begin the infusion. Check for drops in the drip chamber.

To Change the Rate While Infusing

1.Enter a new rate.

2.Press OK.

NOTE: If the OK key is not pressed within 10 seconds of the change, the infusion will continue at the old rate.

Vistabasic Instructions for Use

9/25

950787 Rev H 02/03

 

Image 10
Contents Vista basic Instructions for UseVista basic Instructions for Use 950787 Rev H 02/03 Vista basic Contents Vista basic Instructions for Use 950787 Rev H 02/03 Safety Standard Introduction Door Open Button Aluminum Case Function KeysTurn on the Pump Preparing the InfusionPrepare the Pump Load the SetRate and Vtbd Volume To Be Delivered Continuous ModeContinuous Mode YESRate and Time To Change the Rate While InfusingVtbd and Time To Change the RateKeep Vein Open Rate Information KEYPiggyback OptionsPiggyback To Review or Change Ramp/Taper Mode Data Piggyback, Ramp/Taper ModeRAMP/TAPER Mode To Exit PiggybackTo Exit Ramp/Taper Mode Ramp/Taper ModeTo Clear Ramp/Taper Mode Fast TaperTo Review or Change Program Mode data Program ModeProgram Mode To Clear Program ModeDelay a Mode Delay a ModeTo Exit Delay a Mode To Review Data During Infusion Dosage CalculationDosage Calculation To Exit Dosage CalculationBolus Without Entering a Volume Bolus ModeInterval Bolus Intermittent Mode Standby Total InfusedTo Clear the Total Infused To Exit Standby FunctionDrop Sensor Battery CapacityOcclusion Pressure To Turn On the Drop SensorTo Turn Off Data Lock Data Lock, Contrast SelectionData Lock Contrast SelectionAdjust Clock Adjust Clock, LoudnessLoudness Alarms Alarm Possible Cause ActionAC and 12 AC, 12 V, BATTERY, & CC InterfaceBattery Operation CC InterfaceService MAINTENANCE, STORAGE, & ServiceMaintenance StorageSpecifications 2001 B. Braun Medical Inc