Braun 950787 Occlusion Pressure, To Turn On the Drop Sensor, To Turn Off the Drop Sensor

Page 20

OPTIONS Occlusion Pressure, Drop Sensor, Battery

OCCLUSION PRESSURE

The pressure at which a downstream occlusion alarm will be triggered may be varied. The higher the pressure limit, the less sensitive the pump is to changes in fluid resistance.

1.Press OPTS until display shows OCCLUSION PRESSURE.

2.Press (--) to change pressure to Low. Press (+) to change pressure to High.

3.Press END to return to Continuous Mode.

NOTE: Low corresponds to approximately 8 PSI and high to approximately 17 PSI. High pressure limits may increase the severity of an infiltration without alarming.

DROP SENSOR

Using a drop sensor eliminates the need for pre-setting a VTBD in Continuous Mode and senses upstream occlusions and empty solution containers.

NOTE: Turning the Drop Sensor feature off forces volumes to be entered and prevents sensing of upstream occlusions.

To Turn On the Drop Sensor

1.Press OPTS until display shows DROP SENSOR.

2.Press ON.

3.Press END to return to Continuous Mode.

To Turn Off the Drop Sensor

1.Press OPTS until display shows DROP SENSOR.

2.Press OFF.

3.Press END to return to Continuous Mode.

BATTERY CAPACITY

This is an indicator of the remaining battery time in hours and minutes. This screen also indicates the amount of time the pump has been in operation.

1.Press OPTS until display shows Battery Left =.

2.Press END to return to Continuous Mode

Vistabasic Instructions for Use

19/25

950787 Rev H 02/03

 

Image 20
Contents Vista basic Instructions for UseVista basic Instructions for Use 950787 Rev H 02/03 Vista basic Contents Vista basic Instructions for Use 950787 Rev H 02/03 Safety Standard Introduction Door Open Button Aluminum Case Function KeysPreparing the Infusion Prepare the PumpTurn on the Pump Load the SetContinuous Mode Continuous ModeRate and Vtbd Volume To Be Delivered YESTo Change the Rate While Infusing Vtbd and TimeRate and Time To Change the RateKeep Vein Open Rate Information KEYPiggyback OptionsPiggyback Piggyback, Ramp/Taper Mode RAMP/TAPER ModeTo Review or Change Ramp/Taper Mode Data To Exit PiggybackRamp/Taper Mode To Clear Ramp/Taper ModeTo Exit Ramp/Taper Mode Fast TaperProgram Mode Program ModeTo Review or Change Program Mode data To Clear Program ModeTo Exit Delay a Mode Delay a ModeDelay a Mode Dosage Calculation Dosage CalculationTo Review Data During Infusion To Exit Dosage CalculationInterval Bolus Intermittent Mode Bolus ModeBolus Without Entering a Volume Total Infused To Clear the Total InfusedStandby To Exit Standby FunctionBattery Capacity Occlusion PressureDrop Sensor To Turn On the Drop SensorData Lock, Contrast Selection Data LockTo Turn Off Data Lock Contrast SelectionLoudness Adjust Clock, LoudnessAdjust Clock Alarms Alarm Possible Cause ActionAC, 12 V, BATTERY, & CC Interface Battery OperationAC and 12 CC InterfaceMAINTENANCE, STORAGE, & Service MaintenanceService StorageSpecifications 2001 B. Braun Medical Inc