MAINTENANCE, STORAGE, & SERVICE

Maintenance

Clean the pump with a soft, lint-free cloth or swab dampened with soap and water, a general non staining chemical disinfectant, isopropyl alcohol (15%), derivatives of phenylphenol (5%) or ammonium chloride (10%).

As a general recommendation for cleaning and disinfecting this device, use a lint-free cloth dampened (not soaked) with a 0.5% bleach (Sodium Hypochlorite) solution (10% solution with water). The solution is most effective when prepared weekly and allowed to remain on the device for approximately 10 minutes prior to wiping off.

Do not use acetone solutions containing glutaraldehyde or abrasive cleansers on the pump. If necessary, Betadine or Iodine solutions may be used, but they will stain the pump.

Do not use abrasive cleaning tools or pressurized spraying devices as these may damage the device.

Do not sterilize the pump using ethylene oxide (EtO) gas.

Do not steam autoclave or immerse the pump in any fluids or cleaning solutions.

Always turn off the pump and disconnect it from the electrical outlet prior to cleaning.

Vistabasic Instructions for Use 950787 Rev H 02/03

Storage

Store the pump away from excessive heat, cold, or humidity. Keep devices plugged into electrical power during storage.

Service

If the pump fails to respond to the operating or troubleshooting procedures listed in this manual and the cause cannot be determined, discontinue use and forward it to an authorized B. Braun Service Center.

Should it be necessary to return a pump for repair, contact Technical Support at

B.Braun Customer Service at 800-627- PUMP. A Returned Materials Authorization number will be provided. Carefully pack the pump (preferably in the original packing) and ship it prepaid to:

B. Braun Medical Inc.

Attn: Service Manager

1601 Wallace Drive

Suite 150

Carrollton, TX 75006 USA

B. Braun cannot assume responsibility for loss or damage to returned pumps while they are in transit.

Product complaints may be sent to the Quality Manager at the above address. With each complaint, please include pump serial number and full description of difficulty, including all settings, types of fluids, times, and alarm messages. Return set used if possible. Contact Technical Support for an RMA number prior to return.

24/25

Image 25
Contents Instructions for Use Vista basicVista basic Instructions for Use 950787 Rev H 02/03 Vista basic Contents Vista basic Instructions for Use 950787 Rev H 02/03 Safety Standard Introduction Function Keys Door Open Button Aluminum CasePrepare the Pump Preparing the InfusionTurn on the Pump Load the SetContinuous Mode Continuous ModeRate and Vtbd Volume To Be Delivered YESVtbd and Time To Change the Rate While InfusingRate and Time To Change the RateInformation KEY Keep Vein Open RatePiggyback OptionsPiggyback RAMP/TAPER Mode Piggyback, Ramp/Taper ModeTo Review or Change Ramp/Taper Mode Data To Exit PiggybackTo Clear Ramp/Taper Mode Ramp/Taper ModeTo Exit Ramp/Taper Mode Fast TaperProgram Mode Program ModeTo Review or Change Program Mode data To Clear Program ModeDelay a Mode Delay a ModeTo Exit Delay a Mode Dosage Calculation Dosage CalculationTo Review Data During Infusion To Exit Dosage CalculationBolus Without Entering a Volume Bolus ModeInterval Bolus Intermittent Mode To Clear the Total Infused Total InfusedStandby To Exit Standby FunctionOcclusion Pressure Battery CapacityDrop Sensor To Turn On the Drop SensorData Lock Data Lock, Contrast SelectionTo Turn Off Data Lock Contrast SelectionAdjust Clock Adjust Clock, LoudnessLoudness Alarm Possible Cause Action AlarmsBattery Operation AC, 12 V, BATTERY, & CC InterfaceAC and 12 CC InterfaceMaintenance MAINTENANCE, STORAGE, & ServiceService StorageSpecifications 2001 B. Braun Medical Inc