Braun 950787 manual To Exit Delay a Mode

Page 16

OPTIONS

Delay A Mode

 

 

 

 

DELAY A MODE

The start of Continuous, Piggyback, Ramp/Taper, and Program Modes may all be delayed. The Vistabasic also provides an option to extend the Infusion Complete alarm at the end of the infusion.

1.Program the infusion.

2.Press OPTS until display shows DELAY A MODE.

3.Press ON.

4.Enter in how many hours and minutes the infusion should begin. NOTE: This is NOT a clock time like 12:30 pm.

5.Press OK.

6.Enter how long, in hours and minutes, after the end of the infusion the pump should wait before alarming.

7.Press OK.

8.Press YES if the pump should run at KVO during the delay times. Press NO if the pump should not infuse during the delay times.

9.Review the data and press OK. To change the data, press EXIT and begin with #3 above.

10. Press

to begin the delayed infusion.

To Exit Delay A Mode

1.Stop the infusion.

2.Press OPTS until display shows DELAY A MODE.

3.Press OFF.

4.Press END to return to the previously active mode.

Vistabasic Instructions for Use

15/25

950787 Rev H 02/03

 

Image 16
Contents Vista basic Instructions for UseVista basic Instructions for Use 950787 Rev H 02/03 Vista basic Contents Vista basic Instructions for Use 950787 Rev H 02/03 Safety Standard Introduction Door Open Button Aluminum Case Function KeysPreparing the Infusion Prepare the PumpTurn on the Pump Load the SetContinuous Mode Continuous ModeRate and Vtbd Volume To Be Delivered YESTo Change the Rate While Infusing Vtbd and TimeRate and Time To Change the RateKeep Vein Open Rate Information KEYPiggyback OptionsPiggyback Piggyback, Ramp/Taper Mode RAMP/TAPER ModeTo Review or Change Ramp/Taper Mode Data To Exit PiggybackRamp/Taper Mode To Clear Ramp/Taper ModeTo Exit Ramp/Taper Mode Fast TaperProgram Mode Program ModeTo Review or Change Program Mode data To Clear Program ModeDelay a Mode Delay a ModeTo Exit Delay a Mode Dosage Calculation Dosage CalculationTo Review Data During Infusion To Exit Dosage CalculationBolus Without Entering a Volume Bolus ModeInterval Bolus Intermittent Mode Total Infused To Clear the Total InfusedStandby To Exit Standby FunctionBattery Capacity Occlusion PressureDrop Sensor To Turn On the Drop SensorData Lock, Contrast Selection Data LockTo Turn Off Data Lock Contrast SelectionAdjust Clock Adjust Clock, LoudnessLoudness Alarms Alarm Possible Cause ActionAC, 12 V, BATTERY, & CC Interface Battery OperationAC and 12 CC InterfaceMAINTENANCE, STORAGE, & Service MaintenanceService StorageSpecifications 2001 B. Braun Medical Inc