Braun 950787 manual To Review Data During Infusion, To Exit Dosage Calculation

Page 17

OPTIONS

Dosage Calculation

 

 

 

 

DOSAGE CALCULATION

This feature allows a rate to be calculated given the concentration and dose.

1.Press OPTS until display shows DOSAGE CALC.

2.Press ON.

3.If concentration unit desired is other than the one displayed, press button under the mcg? until units desired appear in the CONCENTRAT. (0*/ml) line.

4.Enter the concentration of the solution. If the concentration of the solution is not known, go to #6 below. Press C to clear current data and re-enter new data.

5.Press OK.

6.To calculate the concentration of the solution, use to move the * to the mg field. Enter the amount of drug in the solution. Then press to move the * to the ml field. Enter the amount of solution.

7.Press OK.

8.Enter the body weight if part of the dosing parameters, then press OK. If body weight is not part of the dosing parameters, press OK with 0 in the body weight field.

9.If time unit desired is other than the one displayed, press button under the /h? until time unit desired appears in the first line. Enter the dose or use Æ to move the * to the rate field and enter the rate.

10.Press OK.

11. Press

to begin the infusion.

To Review Data During Infusion

1. Press INFO twice.

To Exit Dosage Calculation

1.Stop the infusion.

2.Press OPTS until display shows DOSAGE CALC.

3.Press OFF.

4.Press END.

Vistabasic Instructions for Use

16/25

950787 Rev H 02/03

 

Image 17
Contents Instructions for Use Vista basicVista basic Instructions for Use 950787 Rev H 02/03 Vista basic Contents Vista basic Instructions for Use 950787 Rev H 02/03 Safety Standard Introduction Function Keys Door Open Button Aluminum CasePrepare the Pump Preparing the InfusionTurn on the Pump Load the SetContinuous Mode Continuous ModeRate and Vtbd Volume To Be Delivered YESVtbd and Time To Change the Rate While InfusingRate and Time To Change the RateInformation KEY Keep Vein Open RatePiggyback OptionsPiggyback RAMP/TAPER Mode Piggyback, Ramp/Taper ModeTo Review or Change Ramp/Taper Mode Data To Exit PiggybackTo Clear Ramp/Taper Mode Ramp/Taper ModeTo Exit Ramp/Taper Mode Fast TaperProgram Mode Program ModeTo Review or Change Program Mode data To Clear Program ModeTo Exit Delay a Mode Delay a ModeDelay a Mode Dosage Calculation Dosage CalculationTo Review Data During Infusion To Exit Dosage CalculationInterval Bolus Intermittent Mode Bolus ModeBolus Without Entering a Volume To Clear the Total Infused Total InfusedStandby To Exit Standby FunctionOcclusion Pressure Battery CapacityDrop Sensor To Turn On the Drop SensorData Lock Data Lock, Contrast SelectionTo Turn Off Data Lock Contrast SelectionLoudness Adjust Clock, LoudnessAdjust Clock Alarm Possible Cause Action AlarmsBattery Operation AC, 12 V, BATTERY, & CC InterfaceAC and 12 CC InterfaceMaintenance MAINTENANCE, STORAGE, & ServiceService StorageSpecifications 2001 B. Braun Medical Inc