AC, 12 V, BATTERY, & CC INTERFACE
AC and 12 V
♦Check AC voltage as per rating label.
♦Insert/screw on power lead (12 V lead in ambulance car) on back.
Battery Operation
♦Switching to internal battery is automatic in the event of power failure or if 12 V or AC is not connected.
♦Charge battery on initial use, following a battery alarm, or after 2 months of
♦Charging time: approx. 16 hours. Longer charging has no detrimental effect.
♦Full battery provides operation for 3.5 hours at 125 mL/hr.
♦Battery capacity is approx. 50% after 2 years.
♦The service life of the battery can be extended by completely discharging it from time to time and then fully recharging it on AC power.
CC Interface
Interface descriptions available from B. Braun. Connection to interface input uses MFC plug.
Documentation
♦All operating data of the Vista™ basic can be called up and logged via external computer
Vista™ basic Instructions for Use | 23/25 |
950787 Rev H 02/03 |
|