Braun 950787 manual To Review or Change Program Mode data, To Clear Program Mode

Page 15

OPTIONS

Program Mode

 

 

 

 

PROGRAM MODE

Program Mode is for infusions requiring up to 9 separate rates and volumes to be delivered one after another. This is designed to deliver infusions for stepped oncology drugs, circadian medication delivery, and other custom programs.

1.Press OPTS until display shows PROGRAM MODE.

2.Press ON.

3.Enter the rate of the first Step.

4.Press OK.

5.Enter the volume of the first Step.

6.Press OK.

7.Continue to enter the rates and volumes for Steps 2-9.

8.If less than 9 Steps are needed, press OK with a rate of zero.

9.Press OK to accept the total Program Volume.

NOTE: Pressing CHNG on this screen returns to Step 1 for review and changes.

10. Press

to start the infusion. Check for drops in the drip chamber.

To Review or Change Program Mode data

1.Press OPTS until display shows PROGRAM MODE.

Note: If Program Mode is infusing, stop the infusion and press EXIT.

2.Press ON.

3.Review the data on each screen and press OK if data is correct or press to clear the current data and re-enter new data. Press OK to accept

new data.

4.OR, press CHNG on Accept Program Volume screen (see #9 NOTE above).

To Clear Program Mode

1.Press OPTS until display shows PROGRAM MODE.

Note: If Program Mode is infusing, stop the infusion and press EXIT.

2.Press CLR.

3.Press YES to clear all the data in Program Mode.

4.Press ON and continue to program a new Program.

To Exit Program Mode

 

1. When infusion is complete, press

to stop the KVO.

2.Press EXIT.

3.Press END to return to Continuous Mode.

Vistabasic Instructions for Use

14/25

950787 Rev H 02/03

 

Image 15
Contents Instructions for Use Vista basicVista basic Instructions for Use 950787 Rev H 02/03 Vista basic Contents Vista basic Instructions for Use 950787 Rev H 02/03 Safety Standard Introduction Function Keys Door Open Button Aluminum CaseLoad the Set Preparing the InfusionPrepare the Pump Turn on the PumpYES Continuous ModeContinuous Mode Rate and Vtbd Volume To Be DeliveredTo Change the Rate To Change the Rate While InfusingVtbd and Time Rate and TimeInformation KEY Keep Vein Open RateOptions PiggybackPiggyback To Exit Piggyback Piggyback, Ramp/Taper ModeRAMP/TAPER Mode To Review or Change Ramp/Taper Mode DataFast Taper Ramp/Taper ModeTo Clear Ramp/Taper Mode To Exit Ramp/Taper ModeTo Clear Program Mode Program ModeProgram Mode To Review or Change Program Mode dataDelay a Mode Delay a ModeTo Exit Delay a Mode To Exit Dosage Calculation Dosage CalculationDosage Calculation To Review Data During InfusionBolus Mode Bolus Without Entering a VolumeInterval Bolus Intermittent Mode To Exit Standby Function Total InfusedTo Clear the Total Infused StandbyTo Turn On the Drop Sensor Battery CapacityOcclusion Pressure Drop SensorContrast Selection Data Lock, Contrast SelectionData Lock To Turn Off Data LockAdjust Clock, Loudness Adjust ClockLoudness Alarm Possible Cause Action AlarmsCC Interface AC, 12 V, BATTERY, & CC InterfaceBattery Operation AC and 12Storage MAINTENANCE, STORAGE, & ServiceMaintenance ServiceSpecifications 2001 B. Braun Medical Inc