Braun 950787 manual Bolus Mode, Bolus Without Entering a Volume, Interval Bolus Intermittent Mode

Page 18

OPTIONS Bolus Mode, Interval Bolus (Intermittent Mode)

BOLUS MODE

This feature allows a bolus to be given during a Continuous Mode infusion.

1.Press OPTS until display shows BOLUS MODE.

2.Press ON.

3.Press RATE.

4.Enter the rate at which the bolus should infuse, then press OK.

5.Press END.

6.To give a bolus during the infusion, press BOL.

7.Enter the amount of bolus to be given in ml.

8.Press OK.

Bolus Without Entering a Volume

1.During the infusion, press BOL and hold down until a second BOL is displayed to the right.

2.Continue to hold down the first BOL while also pressing the second BOL.

A bolus will be given as long as both BOL keys are held down. The pump will display the amount of bolus being administered and will beep once after each milliliter delivered.

INTERVAL BOLUS (INTERMITTENT MODE)

The basal/KVO rate and entire bag volume must be entered in CONTINUOUS MODE first.

1.Press OPTS until display shows BOLUS MODE.

2.Press ON.

3.Press TIME.

4.Enter the time between the start of each dose, then press OK.

5.Press RATE.

6.Enter the rate at which the dose is to be given, then press OK.

7.Press VTBD.

8.Enter the volume of each dose, then press OK.

9.Press END.

10. Press

. The pump will deliver at the basal/KVO rate for the interval

time, then begin delivery of the first dose.

 

Vistabasic Instructions for Use

17/25

950787 Rev H 02/03

 

 

Image 18
Contents Vista basic Instructions for UseVista basic Instructions for Use 950787 Rev H 02/03 Vista basic Contents Vista basic Instructions for Use 950787 Rev H 02/03 Safety Standard Introduction Door Open Button Aluminum Case Function KeysTurn on the Pump Preparing the InfusionPrepare the Pump Load the SetRate and Vtbd Volume To Be Delivered Continuous ModeContinuous Mode YESRate and Time To Change the Rate While InfusingVtbd and Time To Change the RateKeep Vein Open Rate Information KEYOptions PiggybackPiggyback To Review or Change Ramp/Taper Mode Data Piggyback, Ramp/Taper ModeRAMP/TAPER Mode To Exit PiggybackTo Exit Ramp/Taper Mode Ramp/Taper ModeTo Clear Ramp/Taper Mode Fast TaperTo Review or Change Program Mode data Program ModeProgram Mode To Clear Program ModeDelay a Mode Delay a ModeTo Exit Delay a Mode To Review Data During Infusion Dosage CalculationDosage Calculation To Exit Dosage CalculationBolus Mode Bolus Without Entering a VolumeInterval Bolus Intermittent Mode Standby Total InfusedTo Clear the Total Infused To Exit Standby FunctionDrop Sensor Battery CapacityOcclusion Pressure To Turn On the Drop SensorTo Turn Off Data Lock Data Lock, Contrast SelectionData Lock Contrast SelectionAdjust Clock, Loudness Adjust ClockLoudness Alarms Alarm Possible Cause ActionAC and 12 AC, 12 V, BATTERY, & CC InterfaceBattery Operation CC InterfaceService MAINTENANCE, STORAGE, & ServiceMaintenance StorageSpecifications 2001 B. Braun Medical Inc