Braun 950787 manual Continuous Mode, Yes, Rate and Vtbd Volume To Be Delivered

Page 9

Continuous Mode

When the pump is turned on, it may display: Tubing Type VISTA Pump Set Confirm you are using a Vista™ Pump Set, then press YES.

The pump may then display: Is This a New Therapy?

YES

Begins in Continuous Mode with all data cleared.

The pump may then display: Prime Set?

YES

Press YES to begin priming set.

Press STOP to end priming.

Press EXIT to begin Continuous Mode.

NO

Begins in Continuous Mode with all data cleared.

Press the green key below your answer.

NO

Returns to the beginning of the last program with all data retained.

YES

Press YES to begin priming set.

Press STOP to end priming.

Press EXIT to return to the beginning of the last program with all data retained.

NO

Returns to the beginning of the last program with all data retained.

CONTINUOUS MODE

This section explains how to program a Continuous Mode infusion. For directions on how to program other Modes and Options, refer to those sections later in this manual.

There are three ways to enter infusion data in Continuous Mode:

RATE and VTBD (Volume To Be Delivered)

1.Enter rate in ml/hr.

2.Press VTBD (Volume To Be Delivered).

3.Enter bag volume in ml.

4.Press OK. (To leave screen without saving changes, press EXIT.)

5. Press

to begin the infusion. Check for drops in the drip

Vistabasic Instructions for Use

8/25

950787 Rev H 02/03

 

Image 9
Contents Instructions for Use Vista basicVista basic Instructions for Use 950787 Rev H 02/03 Vista basic Contents Vista basic Instructions for Use 950787 Rev H 02/03 Safety Standard Introduction Function Keys Door Open Button Aluminum CasePrepare the Pump Preparing the InfusionTurn on the Pump Load the SetContinuous Mode Continuous ModeRate and Vtbd Volume To Be Delivered YESVtbd and Time To Change the Rate While InfusingRate and Time To Change the RateInformation KEY Keep Vein Open RateOptions PiggybackPiggyback RAMP/TAPER Mode Piggyback, Ramp/Taper ModeTo Review or Change Ramp/Taper Mode Data To Exit PiggybackTo Clear Ramp/Taper Mode Ramp/Taper ModeTo Exit Ramp/Taper Mode Fast TaperProgram Mode Program ModeTo Review or Change Program Mode data To Clear Program ModeDelay a Mode Delay a ModeTo Exit Delay a Mode Dosage Calculation Dosage CalculationTo Review Data During Infusion To Exit Dosage CalculationBolus Mode Bolus Without Entering a VolumeInterval Bolus Intermittent Mode To Clear the Total Infused Total InfusedStandby To Exit Standby FunctionOcclusion Pressure Battery CapacityDrop Sensor To Turn On the Drop SensorData Lock Data Lock, Contrast SelectionTo Turn Off Data Lock Contrast SelectionAdjust Clock, Loudness Adjust ClockLoudness Alarm Possible Cause Action AlarmsBattery Operation AC, 12 V, BATTERY, & CC InterfaceAC and 12 CC InterfaceMaintenance MAINTENANCE, STORAGE, & ServiceService StorageSpecifications 2001 B. Braun Medical Inc