Braun 950787 manual To Clear Ramp/Taper Mode, To Exit Ramp/Taper Mode, Fast Taper

Page 14

OPTIONS

Ramp/Taper Mode

 

 

 

 

To Clear Ramp/Taper Mode

1.Press OPTS until display shows RAMP TAPER MODE.

2.Press CLR.

3.Press YES to clear all the data in Ramp/Taper Mode.

4.Press ON and continue to program a new Ramp/Taper program.

To Exit Ramp/Taper Mode

 

1. When infusion is complete, press

to stop the KVO.

2.Press EXIT.

3.Press END to return to Continuous Mode.

Fast Taper

The infusion can be interrupted to begin an early taper down process during the Ramp/Taper Mode Infusion.

1.Stop the infusion.

2.Press FAST.

3.Enter the FAST Taper Time in hours and minutes.

4.Press OK.

5. Press

to start the infusion. Check for drops in the drip chamber.

Vistabasic Instructions for Use

13/25

950787 Rev H 02/03

 

Image 14
Contents Vista basic Instructions for UseVista basic Instructions for Use 950787 Rev H 02/03 Vista basic Contents Vista basic Instructions for Use 950787 Rev H 02/03 Safety Standard Introduction Door Open Button Aluminum Case Function KeysTurn on the Pump Preparing the InfusionPrepare the Pump Load the SetRate and Vtbd Volume To Be Delivered Continuous ModeContinuous Mode YESRate and Time To Change the Rate While InfusingVtbd and Time To Change the RateKeep Vein Open Rate Information KEYPiggyback OptionsPiggyback To Review or Change Ramp/Taper Mode Data Piggyback, Ramp/Taper ModeRAMP/TAPER Mode To Exit PiggybackTo Exit Ramp/Taper Mode Ramp/Taper ModeTo Clear Ramp/Taper Mode Fast TaperTo Review or Change Program Mode data Program ModeProgram Mode To Clear Program ModeTo Exit Delay a Mode Delay a ModeDelay a Mode To Review Data During Infusion Dosage CalculationDosage Calculation To Exit Dosage CalculationInterval Bolus Intermittent Mode Bolus ModeBolus Without Entering a Volume Standby Total InfusedTo Clear the Total Infused To Exit Standby FunctionDrop Sensor Battery CapacityOcclusion Pressure To Turn On the Drop SensorTo Turn Off Data Lock Data Lock, Contrast SelectionData Lock Contrast SelectionLoudness Adjust Clock, LoudnessAdjust Clock Alarms Alarm Possible Cause ActionAC and 12 AC, 12 V, BATTERY, & CC InterfaceBattery Operation CC InterfaceService MAINTENANCE, STORAGE, & ServiceMaintenance StorageSpecifications 2001 B. Braun Medical Inc