SPECIFICATIONS
Mechanism | Linear peristaltic volumetric infusion pump |
Dimensions WxHxD | 5.5 in (14cm) x 9.4 in (24cm) x 7.8 in (20cm) |
Weight | 6.8 lbs. (3.1 kg) |
Power Requirements | |
Leakage Current | Max 30µA |
Battery | 7.2V, 1.2 Ah NiCd (rechargeable) |
Battery Operating Time | =3.5 hours at 125 mL/hr |
Rate Range | 0.1 to 800.0 mL/hr in 0.1 mL increments with Vista™ |
| tubing |
Volume to be Infused | 0.1 to 9999.9 mL |
Range |
|
Occlusion Pressure | Variable Settings Low: 8 PSI High: 17 PSI |
KVO Rate | 1 mL/hr or selected rate, whichever is lower |
Flow Rate Accuracy | ± 5% measurement time 8hr, temp. 71.6° F |
Alarms | |
| Downstream Occlusion, Door Open, Pump Stopped, |
| KVO, Unit Malfunction |
| Air bubble size 0.3mL* or 1.5mL/hr** |
| *Can be set from 0.01 to 0.3 mL in service |
| program |
| **Can be set from 0.5 to 3.5 mL/hr in service |
| program |
Memory | User option to retain last Continuous and Piggyback |
| Mode data upon power up. Permanent data |
| retention in Program and Ramp/Taper Modes after |
| pump is turned off |
Tubing | B. Braun Vista™ administration sets |
Modes/Features | Continuous and Piggyback Modes, Program Mode |
| for variable infusions, Ramp/Taper Mode for TPN |
| infusions and Delay A Mode for postponing infusion |
| start and finish |
Operating | Operating: 64.4° F to 95° F, 10.15 PSI to 15.37PSI |
Temperatures | |
| 15.37 PSI |
Vista™ basic Instructions for Use | 25/25 |
950787 Rev H 02/03 |
|