Continuous Mode

KEEP VEIN OPEN RATE

After an infusion is complete, the pump will alarm and begin pumping at 1.0 ml/hr or the current infusion rate, whichever is less.

1.Stop the infusion.

2.Turn off the pump or reset the program.

INFORMATION KEY

This key is only available while the pump is infusing.

1.Press INFO.

2.The Total Infused value and the Total Time of the infusion completed are displayed.

3.Press INFO again and the Battery Left and Total Operation time of the pump is displayed.

4.The pump will return to the main infusion screen after 10 seconds.

Vistabasic Instructions for Use

10/25

950787 Rev H 02/03

 

Image 11
Contents Instructions for Use Vista basicVista basic Instructions for Use 950787 Rev H 02/03 Vista basic Contents Vista basic Instructions for Use 950787 Rev H 02/03 Safety Standard Introduction Function Keys Door Open Button Aluminum CaseLoad the Set Preparing the InfusionPrepare the Pump Turn on the PumpYES Continuous ModeContinuous Mode Rate and Vtbd Volume To Be DeliveredTo Change the Rate To Change the Rate While InfusingVtbd and Time Rate and TimeInformation KEY Keep Vein Open RatePiggyback OptionsPiggyback To Exit Piggyback Piggyback, Ramp/Taper ModeRAMP/TAPER Mode To Review or Change Ramp/Taper Mode DataFast Taper Ramp/Taper ModeTo Clear Ramp/Taper Mode To Exit Ramp/Taper ModeTo Clear Program Mode Program ModeProgram Mode To Review or Change Program Mode dataTo Exit Delay a Mode Delay a ModeDelay a Mode To Exit Dosage Calculation Dosage CalculationDosage Calculation To Review Data During InfusionInterval Bolus Intermittent Mode Bolus ModeBolus Without Entering a Volume To Exit Standby Function Total InfusedTo Clear the Total Infused StandbyTo Turn On the Drop Sensor Battery CapacityOcclusion Pressure Drop SensorContrast Selection Data Lock, Contrast SelectionData Lock To Turn Off Data LockLoudness Adjust Clock, LoudnessAdjust Clock Alarm Possible Cause Action AlarmsCC Interface AC, 12 V, BATTERY, & CC InterfaceBattery Operation AC and 12Storage MAINTENANCE, STORAGE, & ServiceMaintenance ServiceSpecifications 2001 B. Braun Medical Inc