OPTIONS

Piggyback

The Vistabasic contains many options to allow custom programming. To access the various options, press OPTS while the pump is stopped. Pressing OPTS each time will move you to the next option. Eventually, the OPTS key will bring you back to the mode from which you began.

Depending on the model, some modes/options may have been disabled and will not appear on the display when the OPTS key is pressed.

PIGGYBACK

The Piggyback feature allows programming of a secondary (Piggyback) infusion. The pump automatically switches from the Piggyback to the Continuous Mode infusion when the Piggyback program is complete. The option is only available if there is data in the Continuous Mode.

1. Connect the Piggyback set to the injection site on the tubing above the pump on the primary set. Open the roller clamp on the Piggyback set.

2. Hang the Piggyback solution container at least 8 inches above the primary solution container.

3.Program the continuous infusion in the Continuous Mode (see “Continuous Mode”).

4.Press OPTS.

5.Press ON to turn the Piggyback feature on.

6. Press END. PIGY will now be displayed in the lower left corner.

7.Press PIGY.

8.Enter the Piggyback infusion data as you would Continuous Mode data.

9. Press

to start the Piggyback infusion.

10. Check that drops are falling in the Piggyback drip chamber.

CAUTION: Do not attempt to infuse both Continuous and Piggyback fluids at the same time. Use primary sets with check valves.

NOTE: To prevent incorrect delivery, the programmed Piggyback volume to be delivered must equal the amount of fluid in the Piggyback container. Automatic Piggyback is not intended for the infusion of fluids requiring flushing before and after administration.

NOTE: If the Piggyback container is not empty, the remaining Piggyback fluid will be delivered at the Continuous Mode rate once the pump transitions back to the Continuous Mode.

Vistabasic Instructions for Use

11/25

950787 Rev H 02/03

 

Image 12
Contents Vista basic Instructions for UseVista basic Instructions for Use 950787 Rev H 02/03 Vista basic Contents Vista basic Instructions for Use 950787 Rev H 02/03 Safety Standard Introduction Door Open Button Aluminum Case Function KeysPreparing the Infusion Prepare the PumpTurn on the Pump Load the SetContinuous Mode Continuous ModeRate and Vtbd Volume To Be Delivered YESTo Change the Rate While Infusing Vtbd and TimeRate and Time To Change the RateKeep Vein Open Rate Information KEYOptions PiggybackPiggyback Piggyback, Ramp/Taper Mode RAMP/TAPER ModeTo Review or Change Ramp/Taper Mode Data To Exit PiggybackRamp/Taper Mode To Clear Ramp/Taper ModeTo Exit Ramp/Taper Mode Fast TaperProgram Mode Program ModeTo Review or Change Program Mode data To Clear Program ModeDelay a Mode Delay a ModeTo Exit Delay a Mode Dosage Calculation Dosage CalculationTo Review Data During Infusion To Exit Dosage CalculationBolus Mode Bolus Without Entering a VolumeInterval Bolus Intermittent Mode Total Infused To Clear the Total InfusedStandby To Exit Standby FunctionBattery Capacity Occlusion PressureDrop Sensor To Turn On the Drop SensorData Lock, Contrast Selection Data LockTo Turn Off Data Lock Contrast SelectionAdjust Clock, Loudness Adjust ClockLoudness Alarms Alarm Possible Cause ActionAC, 12 V, BATTERY, & CC Interface Battery OperationAC and 12 CC InterfaceMAINTENANCE, STORAGE, & Service MaintenanceService StorageSpecifications 2001 B. Braun Medical Inc