Lifescan SureStep Pro manual New Test Memory full Space Data

Page 101

MESSAGE

POSSIBLE

SOLUTION

CAUSE

 

 

 

 

 

No Room for

Bedside unit

Transfer data to free memory

New Test

memory full.

space.

Data

 

 

 

 

 

No Room for

There are five new

Transfer data and validate

New Test

(non-validated)

new test strip lots.

Strip Lot

test strip lots in

 

 

bedside unit list.

 

 

 

 

Phone Line

1. Workstation

1.Workstation modem’s

Busy at

modem phone

phone line must be a

Workstation

line being used

dedicated line. Do not

 

by fax machine.

share line with a fax

 

 

machine.

 

2. Faulty telephone

2.Have phone line inspected.

 

line.

 

 

 

 

QC Due for

Unit will be in QC

Perform required QC tests

Bedside Unit

lockout if required

within 2 hours.

in Less than

QC not performed

 

Two Hours

within 2 hours.

 

 

 

 

QC Due:

QC overdue or QC

Successfully perform all

QC Missed or

test failed.

required QC tests.

Failed

 

 

 

 

 

Reagent Lot

Reagent (solution)

Enter and use new reagent

Expired

lot expired.

lot. Discard any remaining

 

 

expired reagent.

 

 

 

Remove Strip

Not enough

Remove test strip. Check

appears after

sample on test

confirmation dot. If white

you insert

strip or test strip

patches or streaks appear, not

test strip.

inserted too

enough blood was applied for

 

quickly.

an accurate result. Repeat test

 

 

with new test strip.

 

 

When “Insert Strip” appears,

 

 

firmly and more slowly insert

 

 

test strip until it stops.

 

 

 

If you cannot resolve a problem, call LifeScan Healthcare Professional Line:

1 800 524-7226 (USA)

1 888 353-0800 (Canada)

89

Image 101
Contents Bedside Unit Operator’s Guide Page It should not be used for Diagnosis of diabetes 2002, 1996 LifeScan, Inc. All rights reserved Table of Contents Table of Contents Preface About This Guide PrefaceVii Conventions Used in This Guide Help Preface Introduction Introduction Front Bedside UnitTest Strip Holder General Operating Information Main Menu lets you select from six different functions Page Scanning a Barcode SureStepPro Test Strips Frontback Important Test Strip Information Precautions Page Infection Control Quality Control Test Quality Control Test When to Perform a Quality Control TestGlucose Control Solutions Performing a Quality Control Test Page Select the control level you wish to run for example, High Select the test strip lot Applying Control Solution to Test Strip Page Quality Control Test Page If the result is Failed Page Quality Control Test Performing a Patient Test Patient TestPatient Test Page Patient Test Applying Blood Sample to Test Strip Apply the blood to the test strip as follows Syringe Front Page Critical High indicates a result above the range Page Patient Test Transferring Data to the Workstation Data ManagementData Management Page Using the Connection Module Select Transfer Data from the Main Menu SureStepPro 2.0 Users Ethernet Last Data Transfer mm/dd/yy hhmm Failed Bedside Unit Memory If you select Patient If you select QC/Lin/Unkn Press Menu to proceed, or turn off the bedside unit General Care Care and MaintenanceCleaning the Test Strip Holder and Lens Cleaning the Outside of the UnitContact points Lens area Care and Maintenance Changing the Batteries Care and Maintenance Page Adjusting the Screen Contrast Linearity Test Special TestsSpecial Tests Select Lot # Special Tests Page Special Tests Page Special Tests Unknown Solution Test Special Tests HIGH003 Special Tests Page Do not remove the test strip until the countdown is complete In-Service Test Select In-Service Mode from Special Tests menu Troubleshooting Troubleshooting Error Possible Solution Cause Error MessagesError Possible Solution Clean bedside unit Detected a Make sure test strip With water to remove Unexpected Messages and Observations Message Possible Solution Cause Check Data Re-enter your ID using New Test Memory full Space Data Shake vial Bedside unit is Patient Results Bedside Unit Configuration OptionsBedside Unit Bedside Unit Configuration OptionsNot need to enter Operator and IDs Quality Control QC Test Period Frequency of performing a Bedside Unit Specifications SpecificationsBarcode Specifications SpecificationsType BF Equipment Cleaning Agents Index Index Unknown solution test Workstation Page For use with bedside units, versions 1.2