Special Tests
•Touch a drop of linearity solution to the center of the pink test square.
If the white pad becomes completely saturated, you have applied too much linearity solution. Repeat the application with a new test strip.
•Check the confirmation dot on the back of the test strip to verify it has turned completely blue, indicating an adequate amount of solution was applied.
If white patches or streaks are visible, you have not applied enough linearity solution. Repeat the application with a new test strip.
•Wipe the linearity solution vial tip with a
9Firmly insert the test strip all the way into the test strip holder until it stops (the confirmation dot should be facing down).
66