ERROR | POSSIBLE | SOLUTION | |
CAUSE | |||
|
| ||
|
|
| |
Error 3 | The bedside unit | 1.Clean bedside unit, | |
Test Strip | detected a problem | especially contact points in | |
| with the position of | lens area, but | |
| |||
| the test strip. | NOTE: Before cleaning, | |
|
| compare color of test strip | |
|
| confirmation dot to Color | |
|
| Chart on test strip bottle. If | |
|
| glucose level is very high | |
|
| or very low, take appropri- | |
|
| ate steps to monitor | |
|
| patient and provide | |
|
| medical care. | |
|
| • Repeat test with new | |
|
| test strip. | |
|
| • Do not move test strip | |
|
| after you insert it. Remove | |
|
| test strip only when test is | |
|
| complete. | |
|
| 2.If you continue to get any | |
|
| error message, obtain a | |
|
| laboratory result. Take | |
|
| appropriate steps to | |
|
| monitor patient and | |
|
| provide medical care while | |
|
| awaiting laboratory result. | |
|
| 3.Contact LifeScan for help. | |
|
|
|
If you cannot resolve a problem, call LifeScan Healthcare Professional Line:
1 800
1 888
81