Glucose Control Solutions
Use SureStepPro Glucose Control Solutions with the SureStepPro Bedside Unit and SureStepPro Test Strips only.
◆IMPORTANT Control Solution Information
Read the SureStepPro Control Solutions package insert for more information.
•Gently shake each vial of control solution before use.
•Do not use a control solution after the expiration date printed on the vial label.
•Discard any unused portion 3 months after opening. Write the opened date on the vial when you first open it.
•The control ranges printed on the SureStepPro Test Strip bottle label are for SureStepPro Control Solutions only.
•Store the control solution at room temperature below 30°C (86°F). Do not refrigerate or freeze.
•The control solutions contain dye that stains clothing.
•The control solutions are not for human consumption.
•Control solution values can be affected at altitudes of 3000 feet or greater. Read the SureStepPro Glucose Control Solutions package insert for specific information and correction factors.
17