MESSAGE | POSSIBLE | SOLUTION | |
CAUSE | |||
|
| ||
|
|
| |
Data | Less than 100 test | Transfer data to free memory | |
Memory Low | records can be | space. | |
| generated. |
| |
|
|
| |
Data Transfer | Bedside unit/ | Retry data transfer. | |
Error | workstation |
| |
| communication |
| |
| error. |
| |
|
|
| |
Enter | High, low, or | Check ranges. Make sure min | |
Narrower | normal control | value is less than max. | |
Range(s) or | range error. |
| |
Min<Max | Min>Max. |
| |
|
|
| |
Insert Strip | Test strip not fully | Remove test strip. When | |
appears after | inserted into test | “Insert Strip” appears, firmly | |
you insert test | strip holder. | insert test strip until it stops. | |
strip. |
|
| |
|
|
| |
Internal | Faulty internal | Retry operation. If problem | |
Communica- | communications. | persists, do not use bedside | |
tion Error |
| unit. Contact LifeScan. | |
|
|
| |
Internal Error | System Error. | Turn off bedside unit. | |
Power Down |
| Contact LifeScan. | |
|
|
| |
Invalid | Conflicting option | Contact system | |
Configuration | set, or | administrator to review | |
| configuration | configuration settings. | |
| corrupted. | Transfer data. | |
|
|
| |
Invalid | Value | ||
Control | >500 mg/dL was | value. | |
Value | entered. |
| |
|
|
| |
Invalid CTL | Expiration date not | Check CTL code on control | |
Code | correct, or CTL code | solution vial and | |
| format invalid. |
| |
|
|
| |
Invalid | Invalid month or | Check expiration date and | |
Expiration | year. Or, date | ||
Date | entered already |
| |
| expired. |
| |
|
|
|
If you cannot resolve a problem, call LifeScan Healthcare Professional Line: