Troubleshooting
MESSAGE | POSSIBLE | SOLUTION | |
CAUSE | |||
|
| ||
|
|
| |
Check | 1. Bedside unit was | 1.Do not remove bedside | |
Connections | removed from | unit from connection | |
at Bedside | connection | module until link is | |
Unit | module before | established and data are | |
| connection was | transferred. | |
| established. |
| |
| 2. Bad connection | 2.Check connection between | |
| between modem | connection module and | |
| and connection | modem. | |
| module. |
| |
| 3. Connection | 3.Contact LifeScan. | |
| module or |
| |
| modem error. |
| |
|
|
| |
Check | 1. Workstation | 1.Turn on workstation | |
Connections | modem not | modem. | |
at | turned on. |
| |
Workstation | 2. Bad connection | 2.Check modem connection | |
| |||
| between modem | at wall jack. | |
| and wall jack. |
| |
| 3. Workstation | 3.Contact LifeScan. | |
| modem error. |
| |
|
|
| |
Check Phone | 1. Bedside unit | 1.If fax is being used by same | |
Line | modem phone | phone line as connection | |
Connections | line being used | module modem, wait for | |
at Bedside | by fax machine. | fax machine and retry | |
Unit |
| connection. | |
| 2. Bad connection | 2.Check modem connection | |
| between modem | at wall jack. | |
| and wall jack. |
| |
| 3. Faulty phone line. | 3.Inspect phone line. | |
|
|
|
If you cannot resolve a problem, call LifeScan Healthcare Professional Line: