Lifescan SureStep Pro manual Quality Control Test

Page 27

CHAPTER 2

Quality Control Test

Three levels of SureStepPro Glucose Control

Solution are available to check that the

SureStepPro Bedside Unit and SureStepPro

Test Strips are functioning properly.

The SureStepPro bedside units are calibrated at the factory. Although an internal verification is performed each time the bedside unit is turned on, a system calibration (bedside unit and test strip) is necessary. This is achieved at the start of each test when you enter the lot number and code for the current bottle of test strips.

If you are training a new operator on how to use the bedside unit, use the glucose control solutions and perform the testing in the In-Service mode. Refer to “In-Service Test” on page 75, for more information.

The number of control levels and frequency of running a QC test have been programmed by your system administrator. High, Low, and Normal control levels are available—any one control level or a combination of levels may have been selected.

15

Image 27
Contents Bedside Unit Operator’s Guide Page It should not be used for Diagnosis of diabetes 2002, 1996 LifeScan, Inc. All rights reserved Table of Contents Table of Contents Preface About This Guide PrefaceVii Conventions Used in This Guide Help Preface Introduction Introduction Front Bedside UnitTest Strip Holder General Operating Information Main Menu lets you select from six different functions Page Scanning a Barcode SureStepPro Test Strips Frontback Important Test Strip Information Precautions Page Infection Control Quality Control Test Quality Control Test When to Perform a Quality Control TestGlucose Control Solutions Performing a Quality Control Test Page Select the control level you wish to run for example, High Select the test strip lot Applying Control Solution to Test Strip Page Quality Control Test Page If the result is Failed Page Quality Control Test Performing a Patient Test Patient TestPatient Test Page Patient Test Applying Blood Sample to Test Strip Apply the blood to the test strip as follows Syringe Front Page Critical High indicates a result above the range Page Patient Test Transferring Data to the Workstation Data ManagementData Management Page Using the Connection Module Select Transfer Data from the Main Menu SureStepPro 2.0 Users Ethernet Last Data Transfer mm/dd/yy hhmm Failed Bedside Unit Memory If you select Patient If you select QC/Lin/Unkn Press Menu to proceed, or turn off the bedside unit General Care Care and MaintenanceCleaning the Test Strip Holder and Lens Cleaning the Outside of the UnitContact points Lens area Care and Maintenance Changing the Batteries Care and Maintenance Page Adjusting the Screen Contrast Linearity Test Special TestsSpecial Tests Select Lot # Special Tests Page Special Tests Page Special Tests Unknown Solution Test Special Tests HIGH003 Special Tests Page Do not remove the test strip until the countdown is complete In-Service Test Select In-Service Mode from Special Tests menu Troubleshooting Troubleshooting Error Possible Solution Cause Error MessagesError Possible Solution Clean bedside unit Detected a Make sure test strip With water to remove Unexpected Messages and Observations Message Possible Solution CauseCheck Data Re-enter your ID using New Test Memory full Space Data Shake vial Bedside unit is Patient Results Bedside Unit Configuration OptionsBedside Unit Bedside Unit Configuration OptionsNot need to enter Operator and IDs Quality Control QC Test Period Frequency of performing a Bedside Unit Specifications SpecificationsBarcode Specifications SpecificationsType BF Equipment Cleaning Agents Index Index Unknown solution test Workstation Page For use with bedside units, versions 1.2