Lifescan SureStep Pro manual Shake vial Bedside unit is

Page 102

Troubleshooting

MESSAGE

POSSIBLE

SOLUTION

CAUSE

 

 

 

 

 

Shake vial

Bedside unit is

Gently shake vial. Touch a

and apply

ready for control

drop of solution to center of

(High,

solution

pink test square.

Normal,

application.

• If white pad becomes satu-

Low) control

 

rated, you have applied

 

 

too much solution. Repeat

 

 

application with new strip.

 

 

• If white patches or streaks

 

 

appear on confirmation

 

 

dot, you have not applied

 

 

enough solution. Repeat

 

 

application with new strip.

 

 

 

Strip Lot

Test strip lot

Enter and use new test strip

Expired

expired.

lot. Discard any remaining

 

 

expired test strips.

 

 

 

This QC Test

QC solution level

Do not perform QC test using

Not Required

was not selected as

this level unless you

 

required test by

understand that if test is

Continue?

administrator.

performed and fails and QC

 

 

lockout mode was selected,

 

 

bedside unit will lock out

 

 

users until this test and all

 

 

required QC tests are

 

 

performed and pass.

 

 

 

Upload

Data transfer not

Transfer data.

Overdue

performed within

 

 

specified time.

 

 

 

 

Wrong

Type of reagent

Check reagent type (low,

Reagent

(solution) entered

normal, high, linearity, or

Type

does not match

unknown) and re-enter

 

test type.

information.

 

 

 

If you cannot resolve a problem, call LifeScan Healthcare Professional Line:

1 800 524-7226 (USA)

1 888 353-0800 (Canada)

90

Image 102
Contents Bedside Unit Operator’s Guide Page It should not be used for Diagnosis of diabetes 2002, 1996 LifeScan, Inc. All rights reserved Table of Contents Table of Contents Preface Preface About This GuideVii Conventions Used in This Guide Help Preface Introduction Introduction Bedside Unit FrontTest Strip Holder General Operating Information Main Menu lets you select from six different functions Page Scanning a Barcode SureStepPro Test Strips Frontback Important Test Strip Information Precautions Page Infection Control Quality Control Test When to Perform a Quality Control Test Quality Control TestGlucose Control Solutions Performing a Quality Control Test Page Select the control level you wish to run for example, High Select the test strip lot Applying Control Solution to Test Strip Page Quality Control Test Page If the result is Failed Page Quality Control Test Patient Test Performing a Patient TestPatient Test Page Patient Test Applying Blood Sample to Test Strip Apply the blood to the test strip as follows Syringe Front Page Critical High indicates a result above the range Page Patient Test Data Management Transferring Data to the WorkstationData Management Page Using the Connection Module Select Transfer Data from the Main Menu SureStepPro 2.0 Users Ethernet Last Data Transfer mm/dd/yy hhmm Failed Bedside Unit Memory If you select Patient If you select QC/Lin/Unkn Press Menu to proceed, or turn off the bedside unit Care and Maintenance General CareCleaning the Outside of the Unit Cleaning the Test Strip Holder and LensContact points Lens area Care and Maintenance Changing the Batteries Care and Maintenance Page Adjusting the Screen Contrast Special Tests Linearity TestSpecial Tests Select Lot # Special Tests Page Special Tests Page Special Tests Unknown Solution Test Special Tests HIGH003 Special Tests Page Do not remove the test strip until the countdown is complete In-Service Test Select In-Service Mode from Special Tests menu Troubleshooting Troubleshooting Error Messages Error Possible Solution CauseError Possible Solution Clean bedside unit Detected a Make sure test strip With water to remove Message Possible Solution Cause Unexpected Messages and ObservationsCheck Data Re-enter your ID using New Test Memory full Space Data Shake vial Bedside unit is Bedside Unit Configuration Options Patient ResultsBedside Unit Configuration Options Bedside UnitNot need to enter Operator and IDs Quality Control QC Test Period Frequency of performing a Specifications Bedside Unit SpecificationsSpecifications Barcode SpecificationsType BF Equipment Cleaning Agents Index Index Unknown solution test Workstation Page For use with bedside units, versions 1.2