Lifescan SureStep Pro manual Quality Control

Page 107

Quality Control

Configuration Option

Meaning

 

 

 

 

Glucose units

The units selected for

mg/dL

displaying and reporting

mmol/L

control and blood glucose

 

results.

 

 

Required QC tests

Three glucose controls are

High

available. Any combination

Low

may be selected.

Normal

 

 

 

Control solution

The passing range for each

ranges (high, low,

control solution level has

normal) for specified

been defined. This range

test strip

may be the same as or

 

different from the range

 

found on the test strip

 

bottle label.

 

 

Result Display

Determines how quality

Numeric Value

control test results are

Pass/Fail Only

displayed.

(Pass/Fail available for

 

SureStepPro 2.0 only)

 

 

 

QC Enforcement

An option to ensure QC is

Warning Only

successfully completed as

Bedside Unit

scheduled. Warning Only

lockout

displays a message but

Operator

allows you to perform a

Lockout

patient test. Bedside Unit

 

lockout displays the same

 

message but does not allow

 

you to perform a patient

 

test. Operator Lockout

 

prevents the operator who

 

has not successfully per-

(Time period for

formed QC testing within

Operator Lockout is

the designated time period

24 hours for

(1 to 365 days) from

SureStepPro 1.2)

performing a patient test.

 

 

95

Image 107
Contents Bedside Unit Operator’s Guide Page It should not be used for Diagnosis of diabetes 2002, 1996 LifeScan, Inc. All rights reserved Table of Contents Table of Contents Preface About This Guide PrefaceVii Conventions Used in This Guide Help Preface Introduction Introduction Front Bedside UnitTest Strip Holder General Operating Information Main Menu lets you select from six different functions Page Scanning a Barcode SureStepPro Test Strips Frontback Important Test Strip Information Precautions Page Infection Control Quality Control Test Quality Control Test When to Perform a Quality Control TestGlucose Control Solutions Performing a Quality Control Test Page Select the control level you wish to run for example, High Select the test strip lot Applying Control Solution to Test Strip Page Quality Control Test Page If the result is Failed Page Quality Control Test Performing a Patient Test Patient TestPatient Test Page Patient Test Applying Blood Sample to Test Strip Apply the blood to the test strip as follows Syringe Front Page Critical High indicates a result above the range Page Patient Test Transferring Data to the Workstation Data ManagementData Management Page Using the Connection Module Select Transfer Data from the Main Menu SureStepPro 2.0 Users Ethernet Last Data Transfer mm/dd/yy hhmm Failed Bedside Unit Memory If you select Patient If you select QC/Lin/Unkn Press Menu to proceed, or turn off the bedside unit General Care Care and MaintenanceCleaning the Test Strip Holder and Lens Cleaning the Outside of the UnitContact points Lens area Care and Maintenance Changing the Batteries Care and Maintenance Page Adjusting the Screen Contrast Linearity Test Special TestsSpecial Tests Select Lot # Special Tests Page Special Tests Page Special Tests Unknown Solution Test Special Tests HIGH003 Special Tests Page Do not remove the test strip until the countdown is complete In-Service Test Select In-Service Mode from Special Tests menu Troubleshooting Troubleshooting Error Possible Solution Cause Error MessagesError Possible Solution Clean bedside unit Detected a Make sure test strip With water to remove Unexpected Messages and Observations Message Possible Solution CauseCheck Data Re-enter your ID using New Test Memory full Space Data Shake vial Bedside unit is Patient Results Bedside Unit Configuration OptionsBedside Unit Bedside Unit Configuration OptionsNot need to enter Operator and IDs Quality Control QC Test Period Frequency of performing a Bedside Unit Specifications SpecificationsBarcode Specifications SpecificationsType BF Equipment Cleaning Agents Index Index Unknown solution test Workstation Page For use with bedside units, versions 1.2