Lifescan SureStep Pro manual Special Tests

Page 84

Special Tests

If you are required to enter the unknown solution lot number, a configuration option may have been selected which would allow you to enter a new unknown solution lot number at the bedside unit. If you do not see the lot number in the list and the Enter Unk Lot # button appears on the screen, press the button and manually enter the unknown solution lot number and expiration date.

6Select the test strip lot

number (and code) from the list displayed. The last test strip lot number selected appears at the top of the list. Use the arrows to scroll through the list if necessary.

Alternatively, scan the barcode on the test strip bottle label.

A configuration option may have been selected which would allow you to enter a new test strip lot number at the bedside unit. If you do not see the lot number in the list and the Enter Strip Lot button appears on the screen, press the button and manually enter the test strip lot number and CTL code.

72

Image 84
Contents Bedside Unit Operator’s Guide Page It should not be used for Diagnosis of diabetes 2002, 1996 LifeScan, Inc. All rights reserved Table of Contents Table of Contents Preface Preface About This GuideVii Conventions Used in This Guide Help Preface Introduction Introduction Bedside Unit FrontTest Strip Holder General Operating Information Main Menu lets you select from six different functions Page Scanning a Barcode SureStepPro Test Strips Frontback Important Test Strip Information Precautions Page Infection Control Quality Control Test When to Perform a Quality Control Test Quality Control TestGlucose Control Solutions Performing a Quality Control Test Page Select the control level you wish to run for example, High Select the test strip lot Applying Control Solution to Test Strip Page Quality Control Test Page If the result is Failed Page Quality Control Test Patient Test Performing a Patient TestPatient Test Page Patient Test Applying Blood Sample to Test Strip Apply the blood to the test strip as follows Syringe Front Page Critical High indicates a result above the range Page Patient Test Data Management Transferring Data to the WorkstationData Management Page Using the Connection Module Select Transfer Data from the Main Menu SureStepPro 2.0 Users Ethernet Last Data Transfer mm/dd/yy hhmm Failed Bedside Unit Memory If you select Patient If you select QC/Lin/Unkn Press Menu to proceed, or turn off the bedside unit Care and Maintenance General CareCleaning the Outside of the Unit Cleaning the Test Strip Holder and LensContact points Lens area Care and Maintenance Changing the Batteries Care and Maintenance Page Adjusting the Screen Contrast Special Tests Linearity TestSpecial Tests Select Lot # Special Tests Page Special Tests Page Special Tests Unknown Solution Test Special Tests HIGH003 Special Tests Page Do not remove the test strip until the countdown is complete In-Service Test Select In-Service Mode from Special Tests menu Troubleshooting Troubleshooting Error Messages Error Possible Solution CauseError Possible Solution Clean bedside unit Detected a Make sure test strip With water to remove Message Possible Solution Cause Unexpected Messages and ObservationsCheck Data Re-enter your ID using New Test Memory full Space Data Shake vial Bedside unit is Bedside Unit Configuration Options Patient ResultsBedside Unit Configuration Options Bedside UnitNot need to enter Operator and IDs Quality Control QC Test Period Frequency of performing a Specifications Bedside Unit SpecificationsSpecifications Barcode SpecificationsType BF Equipment Cleaning Agents Index Index Unknown solution test Workstation Page For use with bedside units, versions 1.2