Error Messages
ERROR | POSSIBLE | SOLUTION | |
CAUSE | |||
|
| ||
|
|
| |
Error 1 | The bedside unit | 1.Repeat test with new test | |
Sample | detected a problem | strip and: | |
Application | with the color of | • Apply blood to pink test | |
| the test strip. | square only. | |
| |||
|
| • Make sure confirmation | |
|
| dot turns completely blue. | |
|
| • Insert test strip within | |
|
| 2 min of applying blood. | |
|
| 2. Check that sample hemat- | |
|
| ocrit is within range: adult | |
|
| ||
|
| 3.If you continue to get any | |
|
| error message, obtain a | |
|
| laboratory result, and | |
|
| • Compare color of test | |
|
| strip confirmation dot to | |
|
| Color Chart on test strip | |
|
| bottle to see if glucose | |
|
| level is very high or very | |
|
| low. | |
|
| • Take appropriate steps | |
|
| to monitor patient and | |
|
| provide medical care while | |
|
| awaiting laboratory result. | |
|
| 4.Clean bedside unit and | |
|
| repeat test with new test | |
|
| strip. | |
|
| 5.Contact LifeScan for help. | |
|
|
|
If you cannot resolve a problem, call LifeScan Healthcare Professional Line:
1 800
1 888
79