Clarity manual DataApex INT9 A/D Card will appear, Setup Control Modules list of the dialog

Page 10
System Configuration

Clarity - INT9 A/D converter

Installation

6.Enter the detector names in Name fields for individual channels, set signal units.

Note: A detailed description of this dialog can be found in chapter 3.1 on pg. 14.

7. Press the OK button.

The DataApex INT9 A/D Card will appear in

the Setup Control Modules list of the dialog.

8.Drag and drop the DataApex INT9 A/D Card icon from the Setup Control Modules list on the left side of the System Configuration dialog

to the desired Instrument

tab on the right

side

(or use the button

to do so).

Fig. 4. System Configuration

9. Set the Start input and Ready output numbers for your acquisition card according to the wires being used for synchronization.

Note: The configuration dialog of the INT9 card (INT9 A/D Card Setup) can be displayed any time by double- clicking on its icon or using the Setup button.

10

Image 10
Contents Clarity Clarity Hardware Page Using multiple INT9 cards Using the INT9 card Brief Description INT9 Acquisition cardHardware and software requirements Advantages of the INT9 A/D converterINT9 Card InstallationSignal cables Standard cable for Clarity stationStarting marker cables Connection with ChromatographDigital output cables Connecting Clarity to chromatograph Clarity configuration Select the DataApex INT9 A/D Card and pressSetup Control Modules list of the dialog DataApex INT9 A/D Card will appearSystem Configuration multiple Instruments System Configuration two INT9 cards Device Manager two INT9 cards INT9 A/D Card Setup Using the INT9 cardNumber of Channels Name Inversion of SignalChannel 1 to BipolarMethod Setup Supply FrequencyRange Select DetectorSample Rate Start Restart External Start/StopStart Only Start StopDigital Inputs and Outputs DownTroubleshooting Locate your problemProblems with INT9 driver Board malfunctionHow to check the INT9 driver Error status of the driverManual Installation of the INT9 drivers Go to Start Programs Accessories System Tools System Restore Reinstallation of drivers in Windows Invoke the Start Control Panel Add/Remove Hardware Select Add/Troubleshoot a device and press the Next button Clarity directory. Then press the Next button Data Acquisition non-functional Data Acquisition Simulated Clarity INT9 A/D converter Troubleshooting Tables and specifications Description of the INT9 card connector MaleParameters of digital Inputs and Outputs INT9 CE Conformity Declaration Technical Data Input impedance Sample rate