Clarity INT9 manual Parameters of digital Inputs and Outputs

Page 33

Clarity - INT9 A/D converter

Tables and specifications

5.2Parameters of digital Inputs and Outputs

The card contains eight digital TTL outputs (OUT1-OUT8) with a maximum permissible current of 20 mA at both levels. The first four outputs are also designed as relay contacts (OUT1S1/OUT1S2- OUT4S1/OUT4S2).

The maximum permissible current is 100 mA, the maximum-switched voltage is 100 V and the contact resistance is 150 mΩ. The relay is onstate (closed) when there is a LOW level of a relevant digital output.

The card contains 4 digital inputs IN1 to IN4.

To suppress contact bouncing the digital inputs are equipped with monostable toggles that prolong any change affecting the input to at least 200 ms (to ensure that the station is able to detect it).

A pulse with duration of at least 1 ms is necessary in order to excite the toggles.

33

Image 33
Contents Clarity Clarity Hardware Page Using multiple INT9 cards Using the INT9 card INT9 Acquisition card Brief DescriptionAdvantages of the INT9 A/D converter Hardware and software requirementsSignal cables InstallationINT9 Card Standard cable for Clarity stationConnection with Chromatograph Starting marker cablesDigital output cables Connecting Clarity to chromatograph Select the DataApex INT9 A/D Card and press Clarity configurationDataApex INT9 A/D Card will appear Setup Control Modules list of the dialogSystem Configuration multiple Instruments System Configuration two INT9 cards Device Manager two INT9 cards Using the INT9 card INT9 A/D Card SetupNumber of Channels Channel 1 to Inversion of SignalName BipolarSupply Frequency Method SetupSelect Detector RangeSample Rate Start Only External Start/StopStart Restart Start StopDown Digital Inputs and OutputsLocate your problem TroubleshootingBoard malfunction Problems with INT9 driverError status of the driver How to check the INT9 driverManual Installation of the INT9 drivers Go to Start Programs Accessories System Tools System Restore Reinstallation of drivers in Windows Invoke the Start Control Panel Add/Remove Hardware Select Add/Troubleshoot a device and press the Next button Clarity directory. Then press the Next button Data Acquisition non-functional Data Acquisition Simulated Clarity INT9 A/D converter Troubleshooting Description of the INT9 card connector Male Tables and specificationsParameters of digital Inputs and Outputs INT9 CE Conformity Declaration Input impedance Sample rate Technical Data