Clarity INT9 manual Connecting Clarity to chromatograph

Page 8

Clarity - INT9 A/D converter

Installation

Digital output (Relay)

 

 

Starting cable

Red

 

 

 

 

Shielding

START (REMOTE, …)

Signal cable

Red

 

White

Shielding

NONSYMMETRICAL CONNECTION

1V (100mV, Hi, +, …)

0 (COM, AGND, …)

Digital output (Relay)

 

 

Starting cable

Red

 

 

 

 

Shielding

START (REMOTE, …)

Signal cable

Red

 

White

Shielding

SYMMETRICAL CONNECTION

+(+1V, Hi, …) - (-1V, Lo, …)

COM (0, GND)

Fig. 2. Connecting Clarity to chromatograph

2.3.1Connecting signal cables:

Signal inputs of the A/D converters are symmetrical: + (red), - (white) and analogue ground (copper braiding).

Caution! Shielding must be connected. It serves not only as the shielding but also as the analogue earth against which the measurement takes place. In the case of asymmetrical output of a detector (only two connectors) the shielding must be connected to the white lead! No lead of the signal cable may remain unconnected.

Try to connect to the detector output of the chromatograph with the largest possible signal level, usually indicated as INTEGRATOR (signal approx. 1V). The level of the signal on the output marked RECORDER is usually only approx. 10mV.

For easier alterations of the wiring, we supply a SV8 Terminal board with screw contacts for INT7, INT9, U-PAD,U-PAD2and Net-PADA/D converters.

Note: This SV8 Terminal Board is not suitable for applications with small signals or with high electromagnetic interference. This is because the SV8 Terminal Board uses leads and screw contacts that are not shielded.

8

Image 8
Contents Clarity Clarity Hardware Page Using multiple INT9 cards Using the INT9 card Brief Description INT9 Acquisition card Hardware and software requirements Advantages of the INT9 A/D converterInstallation Signal cablesINT9 Card Standard cable for Clarity stationDigital output cables Connection with ChromatographStarting marker cables Connecting Clarity to chromatograph Clarity configuration Select the DataApex INT9 A/D Card and pressSetup Control Modules list of the dialog DataApex INT9 A/D Card will appearSystem Configuration multiple Instruments System Configuration two INT9 cards Device Manager two INT9 cards Number of Channels Using the INT9 cardINT9 A/D Card Setup Inversion of Signal Channel 1 toName BipolarMethod Setup Supply FrequencySample Rate Select DetectorRange External Start/Stop Start OnlyStart Restart Start StopDigital Inputs and Outputs DownTroubleshooting Locate your problemProblems with INT9 driver Board malfunctionHow to check the INT9 driver Error status of the driverManual Installation of the INT9 drivers Go to Start Programs Accessories System Tools System Restore Reinstallation of drivers in Windows Invoke the Start Control Panel Add/Remove Hardware Select Add/Troubleshoot a device and press the Next button Clarity directory. Then press the Next button Data Acquisition non-functional Data Acquisition Simulated Clarity INT9 A/D converter Troubleshooting Tables and specifications Description of the INT9 card connector MaleParameters of digital Inputs and Outputs INT9 CE Conformity Declaration Technical Data Input impedance Sample rate