Clarity - INT9 A/D converter

Using the INT9 card

Select Detector

Selects the detector whose parameters will be set. If there is only one detector (channel) configured on the Clarity Instrument, this control will not be displayed.

Range

Selects the input voltage range (in mV). The selected range corresponds to the maximum output voltage of the detector that the A/D converter can still process.

Available ranges for INT9 card are 156, 1250 and 12 123 mV.

Sample Rate

Depending on Supply Frequency filter selected in the System Configuration - INT9 A/D Card Setup dialog, the available sampling rates are 6.3, 12.5, 25, 50, 100, 200 and 400 samples per second (for 50Hz mains frequency filter) or 7.5, 15, 30, 60, 120, 240 and 480 samples per second (for 60Hz mains frequency filter).

Higher sampling rate allows for the measurement of narrower peaks, but it also means a larger amount of data which will affect the size of the resulting chromatogram and the speed of its processing.

Sample rate sufficient for successful peak detection is about 20 samples per narrowest peak.

A higher sampling rate also increases the amount of noise.

Note: Range and Sample Rate are set individually for each detector, as opposed to Mains Frequency set for the INT9 A/D card as a whole.

17

Image 17
Contents Clarity Clarity Hardware Page Using multiple INT9 cards Using the INT9 card INT9 Acquisition card Brief DescriptionAdvantages of the INT9 A/D converter Hardware and software requirementsSignal cables InstallationINT9 Card Standard cable for Clarity stationDigital output cables Connection with ChromatographStarting marker cables Connecting Clarity to chromatograph Select the DataApex INT9 A/D Card and press Clarity configurationDataApex INT9 A/D Card will appear Setup Control Modules list of the dialogSystem Configuration multiple Instruments System Configuration two INT9 cards Device Manager two INT9 cards Number of Channels Using the INT9 card INT9 A/D Card Setup Channel 1 to Inversion of SignalName BipolarSupply Frequency Method SetupSample Rate Select DetectorRange Start Only External Start/StopStart Restart Start StopDown Digital Inputs and OutputsLocate your problem TroubleshootingBoard malfunction Problems with INT9 driverError status of the driver How to check the INT9 driverManual Installation of the INT9 drivers Go to Start Programs Accessories System Tools System Restore Reinstallation of drivers in Windows Invoke the Start Control Panel Add/Remove Hardware Select Add/Troubleshoot a device and press the Next button Clarity directory. Then press the Next button Data Acquisition non-functional Data Acquisition Simulated Clarity INT9 A/D converter Troubleshooting Description of the INT9 card connector Male Tables and specificationsParameters of digital Inputs and Outputs INT9 CE Conformity Declaration Input impedance Sample rate Technical Data