Clarity INT9 manual Data Acquisition Simulated

Page 30

Clarity - INT9 A/D converter

Troubleshooting

If your A/D card is not in the left-hand Setup Control Modules list, add it using the Add button and repeat the previous step.

Note: More information on configuration of the INT9 card is in chapter 2.4 - Clarity configuration on pg. 9.

b)You have a licence purchased for collections from a smaller number of instruments:

Solution: Open the System Configuration dialog from the main Clarity window using the System - Configuration command and check the tab of the corresponding instrument – Instrument x. Blue symbol of the curve in the tab header indicates an Instrument not usable for measuring.

Check your serial number S/N (for example) using the Help - About command from the main Clarity window.

c)You are using Clarity Offline or a DEMO version, which does not enable the measurement of chromatograms.

Solution: Check whether there is a blue line with the title OFFLINE displayed in the main Clarity window under the symbols of the instruments, or the title DEMO in the window header.

4.5Data Acquisition - Simulated

The label “Simulated” is displayed in the Data Acquisition window

The corresponding instrument only displays the simulated curve (from the file CHANNX.DTA), or is allocated as what is known as a DEMO driver.

Solution: Open the System Configuration dialog from the main Clarity window using the System - Configuration command and check the tab of the corresponding instrument - Instrument x.

From the list of equipment allocated to the instrument, remove the Detector x from DEMO driver field and draw the correct detector of the A/D card you are using from the list on the left.

30

Image 30
Contents Clarity Clarity Hardware Page Using multiple INT9 cards Using the INT9 card Brief Description INT9 Acquisition cardHardware and software requirements Advantages of the INT9 A/D converterINT9 Card InstallationSignal cables Standard cable for Clarity stationConnection with Chromatograph Starting marker cablesDigital output cables Connecting Clarity to chromatograph Clarity configuration Select the DataApex INT9 A/D Card and pressSetup Control Modules list of the dialog DataApex INT9 A/D Card will appearSystem Configuration multiple Instruments System Configuration two INT9 cards Device Manager two INT9 cards Using the INT9 card INT9 A/D Card SetupNumber of Channels Name Inversion of SignalChannel 1 to BipolarMethod Setup Supply FrequencySelect Detector RangeSample Rate Start Restart External Start/StopStart Only Start StopDigital Inputs and Outputs DownTroubleshooting Locate your problemProblems with INT9 driver Board malfunctionHow to check the INT9 driver Error status of the driverManual Installation of the INT9 drivers Go to Start Programs Accessories System Tools System Restore Reinstallation of drivers in Windows Invoke the Start Control Panel Add/Remove Hardware Select Add/Troubleshoot a device and press the Next button Clarity directory. Then press the Next button Data Acquisition non-functional Data Acquisition Simulated Clarity INT9 A/D converter Troubleshooting Tables and specifications Description of the INT9 card connector MaleParameters of digital Inputs and Outputs INT9 CE Conformity Declaration Technical Data Input impedance Sample rate