Clarity INT9 manual Channel 1 to, Name, Inversion of Signal, Bipolar, Set Units

Page 15

Clarity - INT9 A/D converter

Using the INT9 card

Channel 1 (to 4)

For each channel of the INT9 A/D Converter the name of the signal can be edited in the

Name field and the Set Units... button can be used to change other signal parametres. Effects of these changes can be seen in other fields of the respective Channel 1 (to 4) section.

Name

This field will be used to describe the signal in the Chromatogram window.

Inversion of Signal

Inverts the polarity of the signal from the detector. This function can also be used for easy correction of inversely connected wires.

Bipolar

Both positive and negative voltages will be measured. Unchecking the Bipolar option will improve the resolution of the signal by 1 bit. On the other hand, it will also restrict the measuring of negative signals (i.e. below zero).

Set Units...

Invokes the Detector Units dialog.

Fig. 9. Detector Units

Quantity – Field for entering the quantity of the units.

Units – Field for entering the unit symbol, along with the prefix.

Coefficient – Allows to enter the coefficient between the selected units and mV (default).

Autoprefix – Allows to enter the unit prefix. In graphs, units will be scaled automatically, while in the tables, units with the prefix entered will be used.

15

Image 15
Contents Clarity Clarity Hardware Page Using multiple INT9 cards Using the INT9 card INT9 Acquisition card Brief DescriptionAdvantages of the INT9 A/D converter Hardware and software requirementsStandard cable for Clarity station InstallationSignal cables INT9 CardConnection with Chromatograph Starting marker cablesDigital output cables Connecting Clarity to chromatograph Select the DataApex INT9 A/D Card and press Clarity configurationDataApex INT9 A/D Card will appear Setup Control Modules list of the dialogSystem Configuration multiple Instruments System Configuration two INT9 cards Device Manager two INT9 cards Using the INT9 card INT9 A/D Card SetupNumber of Channels Bipolar Inversion of SignalChannel 1 to NameSupply Frequency Method SetupSelect Detector RangeSample Rate Start Stop External Start/StopStart Only Start RestartDown Digital Inputs and OutputsLocate your problem TroubleshootingBoard malfunction Problems with INT9 driverError status of the driver How to check the INT9 driverManual Installation of the INT9 drivers Go to Start Programs Accessories System Tools System Restore Reinstallation of drivers in Windows Invoke the Start Control Panel Add/Remove Hardware Select Add/Troubleshoot a device and press the Next button Clarity directory. Then press the Next button Data Acquisition non-functional Data Acquisition Simulated Clarity INT9 A/D converter Troubleshooting Description of the INT9 card connector Male Tables and specificationsParameters of digital Inputs and Outputs INT9 CE Conformity Declaration Input impedance Sample rate Technical Data