Clarity - INT9 A/D converter | Using the INT9 card |
3.2.2Method Setup - Measurement
Fig. 11. Method Setup - Measurement
External Start/Stop
Enables control from an external signal.
Note: The Input used for an external Start from a chromatograph can be set in the System Configuration dialog (see Fig. 4 on pg. 10).
Start Only
The external signal only starts an analysis. Triggering the input will not influence the acquisition in progress.
Start - Restart
The external signal terminates the analysis in progress and at the same time starts a new one. The first signal starts the analysis; the second signal terminates it and at the same time starts a new one - the
Start - Stop
The external signal starts and stops an analysis. The first signal starts the analysis; the second signal stops it, etc.