|
| Flue Gas Analysis System | Operation 4 | |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
If, on the other hand, reliable analysis, free of instrument distortion, is required from the very onset of operation, TBE/AI recommends a 24 hour
In either case, emlploy the following procedure:
(1)Set the analyzer RANGE switch to the
(2)Rotate the
Note: When the valve is rotated so that the handle index faces the inside of the case (open port), atmosplleric air is drawn in by the pump and flows through the oxygen analyzer. Be sure that the valve is returned to the sample inlet position after calibration is completed. Never leave the valve in the
(3)Allow air to flow through the unit for 3 to 5 minutes.
(4)Unlock and adjust the SPAN control until the meter pointer is
in coincidence with the CAL mark (20.9% O2) on the meter scale. Relock the SPAN control.
(5)Rotate the selector valve so that sample gas is flowing through the analyzer.
(6)Select the range of analysis that will provide the best possible resolution of the oxygen content of the sample gas.
(7)Allow sample gas, not calibration air, to flow through the analyzer for the duration of the equilibration period.
Note: It is particularly vital to the condition of the
TELEDYNE BROWN ENGINEERING |
Analytical Instruments