Teledyne 9700 instruction manual Flue Gas Analysis System

Page 31

 

 

Flue Gas Analysis System

Operation 4

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

If, on the other hand, reliable analysis, free of instrument distortion, is required from the very onset of operation, TBE/AI recommends a 24 hour run-in period before operational calibration and service .

In either case, emlploy the following procedure:

(1)Set the analyzer RANGE switch to the 0–25% position. Power has now been applied to the instrument circuitry and to the sample pump.

(2)Rotate the sample-caibration selector valve so that calibration or span gas is being delivered to the analyzer.

Note: When the valve is rotated so that the handle index faces the inside of the case (open port), atmosplleric air is drawn in by the pump and flows through the oxygen analyzer. Be sure that the valve is returned to the sample inlet position after calibration is completed. Never leave the valve in the half-way (closed) position with the pump running.

(3)Allow air to flow through the unit for 3 to 5 minutes.

(4)Unlock and adjust the SPAN control until the meter pointer is

in coincidence with the CAL mark (20.9% O2) on the meter scale. Relock the SPAN control.

(5)Rotate the selector valve so that sample gas is flowing through the analyzer.

(6)Select the range of analysis that will provide the best possible resolution of the oxygen content of the sample gas.

(7)Allow sample gas, not calibration air, to flow through the analyzer for the duration of the equilibration period.

Note: It is particularly vital to the condition of the A-3 cell employed in flue gas (or high CO2 content applications) that the cell not be exposed to a CO2 free environment for a prolonged period of time.

TELEDYNE BROWN ENGINEERING

4-3

Analytical Instruments

Image 31
Contents Model Flue Gas Analysis System Important Notice WarrantyTable of Contents Maintenance & Troubleshooting Introduction Flue Gas Analysis SystemOverview Sample Conditioning General InformationSample Conditioning Oxygen Analyzer Section Standard FeaturesOxygen Analyzer Section General Information Optional Features General Information Combustible Gas Analyzer Section Combustible Gas Analyzer SectionGeneral Information Compound LEL Response Factor Operating Controls and Indicators Meter Trim Recessed Secondary ControlsAnalog Output Current MAX Load Output Impedance Oxygen Analyzer Theory of OperationTheory of Operation Combustible Analyzer Detector Flue Gas Analysis System Theory of Operation Electrical Connections InstallationPower Output Signal VoltageInterconnection Diagram Output Signal Current Alarm and/or Control CircuitryInstallation Meter Zeroing OperationStartup of Oxygen Analyzer PreliminaryInitial Calibration and Equilibration Flue Gas Analysis System Operational Calibration Routine Operational Calibration ProcedureStartup of Combustibles Analyzer Operation Sample System Startup Span GasSystem Startup Analyzer StartupOperation Oxygen Cell Replacement Flue Gas Analysis System Maintenance & TroubleshootingMaintenance Routine MaintenanceCell Warranty Combustible Sensor Spray NozzleWater Strainer Filter Condensate TrapPump TroubleshootingGeneral Oxygen Analyzer .2.2.1 Inability to Calibrate Inability to Calibrate Combustible Analyzer SectionNo Alarm With High Gas Level Lamp Failure Appendix SpecificationsOxygen Analysis Section Appendix ModelModel 9700 System Standard Unit Only Recommended Spare Parts ListReference Drawings Response of Combustible Sensor to Various Gases Appendix Model Supporting Equipment for Flue Gas Analysis Systems Figure A-1 Typical Dry Probe Maintenance General Maintenance for CC-2B Series PumpsRepair Appendix Model Flue Gas Analysis System Figure A-3 Pump Assembly Flue Gas Analysis System Figure A-4 Replacing Flex-I-Liner Replacing the Liner in the Flex-I-Liner PumpFlue Gas Analysis System