Teledyne 9700 instruction manual Sample Conditioning

Page 7

 

 

Flue Gas Analysis System

General Information

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Next, the gas and water mixture enter the pump. The pump contains no valves or packing and will handle the dirtiest of samples without any detrimental effect

Exiting the pump under positive pressure, the sample mixture is forced through a mixer which contains a series of baffles which function to thoroughly scrub the sample gas.

At the next stage of conditioning, the sample solution enters a separa- tor where water is gravitated to drainage. The scrubbed flue gas sample then exits the separator through a sample outlet and enters a filter-conden- sate trap for final removal of particulates and water before undergoing analysis.

The sample gas stream, having passed the final filtration trap, flows into a tee. Here, the stream is divided with most of the sample gas being vented to atmosphere, while the balance is routed to the oxygen analyzer section. Total flow rate is approximately 20 to 30 SCFH. A back pressure control valve is adjusted to permit proper flow past the oxygen analyzer sensor where the percentage of oxygen concentration in the flue gas sample is determined and displayed on the oxygen analyzer meter.

Following oxygen analysis, the sample passes to the combustible analysis section.

Figure 2: Sample Conditioning

TELEDYNE BROWN ENGINEERING

3

Analytical Instruments

Image 7
Contents Model Flue Gas Analysis System Important Notice WarrantyTable of Contents Maintenance & Troubleshooting Introduction Flue Gas Analysis SystemOverview Sample Conditioning General InformationSample Conditioning Oxygen Analyzer Section Standard FeaturesOxygen Analyzer Section General Information Optional Features General Information Combustible Gas Analyzer Section Combustible Gas Analyzer SectionGeneral Information Compound LEL Response Factor Operating Controls and Indicators Meter Trim Recessed Secondary ControlsAnalog Output Current MAX Load Output Impedance Oxygen Analyzer Theory of OperationTheory of Operation Combustible Analyzer Detector Flue Gas Analysis System Theory of Operation Electrical Connections InstallationPower Output Signal VoltageInterconnection Diagram Output Signal Current Alarm and/or Control CircuitryInstallation Meter Zeroing OperationStartup of Oxygen Analyzer PreliminaryInitial Calibration and Equilibration Flue Gas Analysis System Operational Calibration Routine Operational Calibration ProcedureStartup of Combustibles Analyzer Operation Sample System Startup Span GasSystem Startup Analyzer StartupOperation Oxygen Cell Replacement Flue Gas Analysis System Maintenance & TroubleshootingMaintenance Routine MaintenanceCell Warranty Combustible Sensor Spray NozzleWater Strainer Filter Condensate TrapPump TroubleshootingGeneral Oxygen Analyzer .2.2.1 Inability to Calibrate Inability to Calibrate Combustible Analyzer SectionNo Alarm With High Gas Level Lamp Failure Appendix SpecificationsOxygen Analysis Section Appendix ModelModel 9700 System Standard Unit Only Recommended Spare Parts ListReference Drawings Response of Combustible Sensor to Various Gases Appendix Model Supporting Equipment for Flue Gas Analysis Systems Figure A-1 Typical Dry Probe Maintenance General Maintenance for CC-2B Series PumpsRepair Appendix Model Flue Gas Analysis System Figure A-3 Pump Assembly Flue Gas Analysis System Figure A-4 Replacing Flex-I-Liner Replacing the Liner in the Flex-I-Liner PumpFlue Gas Analysis System