Teledyne 9700 instruction manual Recessed Secondary Controls, Meter Trim, Analog Output

Page 17

 

 

Flue Gas Analysis System

General Information

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

4.2Recessed Secondary Controls

NOTE: The recessed potentiometric controls are provided for calibration purposes. They should not be changed once calibration is estab- lished.

The SPAN control adjusts for manufacturing variations in sensitivity between elements and for various gases.

The ZERO control adjusts for zero meter reading with zero gas (air) flowing through the sample cell.

The ALARM control allows the point at which the unit goes into alarm to be adjusted anywhere between 0% and 5% combustibles as shown by the meter reading.

(Optional) The CAUTION control allows the point at which the unit goes into alarm to be adjusted anywhere within the range of the instrument as shown by the meter indication.

4.3Meter Trim

The small potentiometer (P5) located on the main circuit board below the meter is used to trim the meter to full scale deflection at the rated output. The potentiometric control is set at the factory.

4.4Analog Output

An analog (voltage or current) signal is provided for remote monitor- ing or recording. The type of signal is a customer option.

Voltage Output

0–1 Volt full scale or less.

NOTE: The remote meter or recorder should have an input impedance greater than 1K-Ohm.

Current Output

One of the following milliampere outputs may be provided:

TELEDYNE BROWN ENGINEERING

13

Analytical Instruments

Image 17
Contents Model Flue Gas Analysis System Important Notice WarrantyTable of Contents Maintenance & Troubleshooting Overview Flue Gas Analysis SystemIntroduction Sample Conditioning General InformationSample Conditioning Oxygen Analyzer Section Standard FeaturesOxygen Analyzer Section General Information Optional Features General Information Combustible Gas Analyzer Section Combustible Gas Analyzer SectionGeneral Information Compound LEL Response Factor Operating Controls and Indicators Analog Output Recessed Secondary ControlsMeter Trim Current MAX Load Output Impedance Oxygen Analyzer Theory of OperationTheory of Operation Combustible Analyzer Detector Flue Gas Analysis System Theory of Operation Power InstallationOutput Signal Voltage Electrical ConnectionsInterconnection Diagram Output Signal Current Alarm and/or Control CircuitryInstallation Startup of Oxygen Analyzer OperationPreliminary Meter ZeroingInitial Calibration and Equilibration Flue Gas Analysis System Operational Calibration Startup of Combustibles Analyzer ProcedureRoutine Operational Calibration Operation System Startup Span GasAnalyzer Startup Sample System StartupOperation Maintenance Flue Gas Analysis System Maintenance & TroubleshootingRoutine Maintenance Oxygen Cell ReplacementCell Warranty Water Strainer Spray NozzleFilter Condensate Trap Combustible SensorGeneral TroubleshootingPump Oxygen Analyzer .2.2.1 Inability to Calibrate Inability to Calibrate Combustible Analyzer SectionNo Alarm With High Gas Level Lamp Failure Appendix SpecificationsOxygen Analysis Section Appendix ModelModel 9700 System Standard Unit Only Recommended Spare Parts ListReference Drawings Response of Combustible Sensor to Various Gases Appendix Model Supporting Equipment for Flue Gas Analysis Systems Figure A-1 Typical Dry Probe Repair General Maintenance for CC-2B Series PumpsMaintenance Appendix Model Flue Gas Analysis System Figure A-3 Pump Assembly Flue Gas Analysis System Figure A-4 Replacing Flex-I-Liner Replacing the Liner in the Flex-I-Liner PumpFlue Gas Analysis System