Teledyne 9700 instruction manual Operation

Page 34

4 Operation

Model 9700

 

 

 

 

(6)After air has been flowing for a couple of minutes, adjust the ZERO control on the control unit (use the small screwdriver supplied) until the meter pointer lines up with the zero mark on the dial (be sure that the mechanical zero of the meter is first adjusted with the power off.)

(7)Turn on supply of span gas and open the SPAN toggle valve. Then adjust the SPAN flow control valve until a flow rate of 2–3 SCFH is indicated on the right hand flowmeter.

Note: Flow rate of span gas should be set to match that of the air as closely as possible. It does not matter what they read as long as they are both the same.

(8)Adjust the SPAN control with the small screwdriver until the meter pointer reads the analyzed value ot the span gas.

(9)If only a very small or no adjustment was required, no further calibration should be needed. If, however, a fairly large adjustment was made, the ZERO adjustment should be rechecked and readjusted, as necessary. If ZERO readjustment was made, check the SPAN adjustment again.

(10)Adjust alarm setpoint to the desired level. The alarm is adjusted by turning ZERO control until the meter reads the desired alarm point. The ALARM level control is then adjusted until the alarm light just comes on. The ZERO control must then be reset to produce a zero reading.

(11)Once the above adjustment has been made, the unit is ready for service.

(12)After calibration is completed, re-establish sample flow and be sure the flow rate is set to match that which was used for span gas.

4-6

TELEDYNE BROWN ENGINEERING

Analytical Instruments

Image 34
Contents Model Flue Gas Analysis System Warranty Important NoticeTable of Contents Maintenance & Troubleshooting Introduction Flue Gas Analysis SystemOverview General Information Sample ConditioningSample Conditioning Standard Features Oxygen Analyzer SectionOxygen Analyzer Section General Information Optional Features General Information Combustible Gas Analyzer Section Combustible Gas Analyzer SectionGeneral Information Compound LEL Response Factor Operating Controls and Indicators Meter Trim Recessed Secondary ControlsAnalog Output Current MAX Load Output Impedance Theory of Operation Oxygen AnalyzerTheory of Operation Combustible Analyzer Detector Flue Gas Analysis System Theory of Operation Output Signal Voltage InstallationPower Electrical ConnectionsInterconnection Diagram Alarm and/or Control Circuitry Output Signal CurrentInstallation Preliminary OperationStartup of Oxygen Analyzer Meter ZeroingInitial Calibration and Equilibration Flue Gas Analysis System Operational Calibration Routine Operational Calibration ProcedureStartup of Combustibles Analyzer Operation Analyzer Startup Span GasSystem Startup Sample System StartupOperation Routine Maintenance Flue Gas Analysis System Maintenance & TroubleshootingMaintenance Oxygen Cell ReplacementCell Warranty Filter Condensate Trap Spray NozzleWater Strainer Combustible SensorPump TroubleshootingGeneral Oxygen Analyzer .2.2.1 Inability to Calibrate Combustible Analyzer Section Inability to CalibrateNo Alarm With High Gas Level Lamp Failure Specifications AppendixAppendix Model Oxygen Analysis SectionRecommended Spare Parts List Model 9700 System Standard Unit OnlyReference Drawings Response of Combustible Sensor to Various Gases Appendix Model Supporting Equipment for Flue Gas Analysis Systems Figure A-1 Typical Dry Probe Maintenance General Maintenance for CC-2B Series PumpsRepair Appendix Model Flue Gas Analysis System Figure A-3 Pump Assembly Flue Gas Analysis System Replacing the Liner in the Flex-I-Liner Pump Figure A-4 Replacing Flex-I-LinerFlue Gas Analysis System