Teledyne 9700 instruction manual No Alarm With High Gas Level

Page 43

 

 

Flue Gas Analysis System

Maintenance & Troubleshooting 5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

original detector is probably defective and requires replacement. If problems still occur in the detector circuit, then the dummy sensor will give a signal that can be tracked through the system.

Measure the voltage at the tiepoint of R7, R8 and R10 with respect to terminal 14 which is the center of the sensor bridge. With the zero potentiometer adjusted up and down, the voltage should change by about 1/4 volt. If this does not occur, then there is probably something wrong in the Wheatstone Bridge circuitry, i. e., the resistors that constitute the bridge along with P1, the zero potentiometer.

Measure the voltage at the output of A1 (A1-1). (The easiest place to check the output voltage of A is at the tiepoint of R15, R30 and C11.) Use the tiepoint of C7 and C8 (oxygen analyzer) for power supply common. The signal should be tracked through the system. With a voltmeter attached to A1–1 move the zero potentiometer up and down. Pin 1 of A1 should reflect this voltage movement by a couple of volts.

With the zero potentiometer set so there is a voltage at A1-1, follow the signal through the circuit, i.e. through to A1-7 and see if the meter deflection is appropriate to the voltage at A1-7. The meter should indicate full scale deflection with 1 V at A1–7. If it does not, replacement of A1 is indicated.

The preceding discussion concerning signal checkout assumes that the power supplies have been verified and are operating properly (see Section 5.2.1.). The regulated power supply should be checked at C8 and C7 for +/- 15 VDC.

The output of the current supply for the detector can be checked at the two test points (located on each side of R6). The voltage should be 0.3 V, which indicates that 300 mA is flowing through the detector.

5.2.3.2 No Alarm With High Gas Level

If there is trouble with the alarm comparator, A3 or transistor Q1, there will be no alarm indication even though the gas level goes above the setpoint. Also, there will be no alarm if the relay coil K2 fails, although relay coil failure is not likely.

TELEDYNE BROWN ENGINEERING

5-7

Analytical Instruments

Image 43
Contents Model Flue Gas Analysis System Important Notice WarrantyTable of Contents Maintenance & Troubleshooting Introduction Flue Gas Analysis SystemOverview Sample Conditioning General InformationSample Conditioning Oxygen Analyzer Section Standard FeaturesOxygen Analyzer Section General Information Optional Features General Information Combustible Gas Analyzer Section Combustible Gas Analyzer SectionGeneral Information Compound LEL Response Factor Operating Controls and Indicators Meter Trim Recessed Secondary ControlsAnalog Output Current MAX Load Output Impedance Oxygen Analyzer Theory of OperationTheory of Operation Combustible Analyzer Detector Flue Gas Analysis System Theory of Operation Electrical Connections InstallationPower Output Signal VoltageInterconnection Diagram Output Signal Current Alarm and/or Control CircuitryInstallation Meter Zeroing OperationStartup of Oxygen Analyzer PreliminaryInitial Calibration and Equilibration Flue Gas Analysis System Operational Calibration Routine Operational Calibration ProcedureStartup of Combustibles Analyzer Operation Sample System Startup Span GasSystem Startup Analyzer StartupOperation Oxygen Cell Replacement Flue Gas Analysis System Maintenance & TroubleshootingMaintenance Routine MaintenanceCell Warranty Combustible Sensor Spray NozzleWater Strainer Filter Condensate TrapPump TroubleshootingGeneral Oxygen Analyzer .2.2.1 Inability to Calibrate Inability to Calibrate Combustible Analyzer SectionNo Alarm With High Gas Level Lamp Failure Appendix SpecificationsOxygen Analysis Section Appendix ModelModel 9700 System Standard Unit Only Recommended Spare Parts ListReference Drawings Response of Combustible Sensor to Various Gases Appendix Model Supporting Equipment for Flue Gas Analysis Systems Figure A-1 Typical Dry Probe Maintenance General Maintenance for CC-2B Series PumpsRepair Appendix Model Flue Gas Analysis System Figure A-3 Pump Assembly Flue Gas Analysis System Figure A-4 Replacing Flex-I-Liner Replacing the Liner in the Flex-I-Liner PumpFlue Gas Analysis System