2620A, 2625A

Users Manual

Instrument Type

Multifunction Calibrator

Table 6-2. Recommended Test Equipment

Minimum Specification

DC Voltage:

Range: 90 mV to 300V dc

Accuracy: .005%

Recommended

Model

Fluke 5700A

AC Voltage:

Frequency

1kHz

100 kHz

Frequency:

Voltage

29 mV to 300V

15 mV to 300V

Accuracy

0.05%

1.25%

Decade Resistance Source

10 kHz

1V rms

 

.0125%

 

 

 

 

 

Ohms

 

 

Accuracy

 

 

 

 

290Ω

 

 

0.0125%

2.9 kΩ

 

 

0.0125%

 

 

29 kΩ

 

 

0.0125%

 

 

290 kΩ

 

 

0.0125%

 

 

2.9 MΩ

 

 

0.0125%

 

 

 

 

 

 

 

Note

General Resistance Inc. Model RDS 66A

Mercury Thermometer

Thermocouple Probe

Room Temperature Oil/Water Bath

The 5700A Calibrator can be used for 0.05% accuracy (rated) on the 3.0 kΩ, 30 kΩ, and 3.0 MΩ ranges. The 5700A can be used for 0.06% accuracy on the 300Ω range.

0.02degrees Celsius resolution Type K

Thermos bottle and cap

Princo ASTM-56C

Fluke P-20K

Multimeter

Signal Generator

Measures +5V dc

Sinewave, 0.5 to 1V rms 10 Hz to 5 kHz

Alternate Equipment List

Fluke 77

Fluke 6011A

(Minimum specifications are the same as in the Standard Equipment List)

Instrument Type

Recommended Model

DC Voltage Calibrator

DMM Calibrator

Function/Signal Generator

Fluke 5440B

Fluke 5100B (for AC Volts only)

Philips PM5193 or Fluke 6011A

6-6

Page 134
Image 134
Fluke user manual 2620A, 2625A, Instrument Type, 2. Recommended Test Equipment Minimum Specification, Recommended Model