Fluke user manual 2620A, 2625A, 8. Command and Query Reference cont

Models: 2625A 2620A

1 197
Download 197 pages 4.31 Kb
Page 114
Image 114

2620A, 2625A

Users Manual

Table 4-8. Command and Query Reference (cont)

TEMP_CONFIG?

TIME

TIME_DATE?

Use the lowest two bits of the value given as individual flags to specify the temperature configuration. Therefore, the value given must be in the range from 0 through 3 or an Execution Error is generated. These settings affect every channel; they cannot be set for each channel individually. If this command is attempted when measurements are active, an Execution Error results.

Bit Number

Value

Meaning

0

0

°C

 

1

°F

1

0

Disable open thermocouple detection

 

1

Enable open thermocouple detection

Temperature Configuration Query

Response to this query corresponds to settings made with the TEMP_CONFIG command, as follows:

Bit Number

Value

Meaning

0

0

°C

 

1

°F

1

0

Disable open thermocouple detection

 

1

Enable open thermocouple detection

Set the instrument clock.

TIME <hours>,<minutes>

<hours> =

(0 .. 23) (24-hour scale, 18:00 = 6:00 pm).

<minutes> =

(0 .. 59)

Invalid values generate an Execution Error. Seconds are automatically set to 00 with the TIME command.

Retrieve Time and Date

Returns comma-separated integer values for time, date, and year as maintained by the instrument clock. These returned values use the TIME and DATE command integer formats as follows:

hours 0-23 minutes 0-59 seconds 0-59

month 1-12

day 1-31

year 00-99

Seconds are returned, but cannot be set other than to 00.

4-44

Page 114