Operating the Instrument from the Front Panel

3

List Button Functions

Activate (Note 1)

N

20

G

D0

Table 3-15. Review Array

Review Points

HJ

LAST

MIN

MAX

LAST

MIN

MAX

 

 

 

Deactivate

(Note 3)

N

or

C

Note 1. Review array can be entered only from a channel that is defined for some type of measurement ("VDC", "VAC", etc.) Review array cannot be entered from a channel that is OFF.

Note 2. All review values (MIN, MAX, LAST for all channels) can be cleared (to "-----") by pressing:

CLEAR KN

This clearing action can be initiated only while viewing the Review values. Any scan in progress is completed before a requested Review clear is carried out.

Review values are cleared when any channel configuration is changed.

Review values are not cleared automatically at the start of a new scan.

Note 3. When returning to Inactive Mode, the instrument returns to the last channel examined in the Review array. This may not be the same channel from which you started viewing the Review array.

Note 4. Alarms are shown with the LAST Review values. See Alarm Indications for a description of alarm annunciation.

All review values for all channels are cleared with any of the following actions:

CLEAR (KN) is selected. This must be done while review data is on the display; otherwise, an error beep results.

Function or number of terminals for any channel is changed.

Range or thermocouple/RTD type for any channel is changed.

The M and/or B scaling value for any channel is changed.

The RTD R0 value for any channel is changed.

The temperature unit is changed.

The measurement rate is changed.

List Button Functions

If the RS-232 interface is active, you can print out the Last values from the Review array (2620A and 2625A) or all values from Data Logger memory (2625A only). This procedure is described in Table 3-16.

In the LASt printout, channels that are defined as OFF are not included. Following is a sample LASt printout:

3-23