Selecting Scan Interval

The scan interval is the period between starts of measurement scans. The resolution of the scan interval is one second; the maximum scan interval is 9 hours: 99 minutes: 99 seconds. The format of the scan interval is x:xx:xx; 1:30:45 translates to 1 hour, 30 minutes and 45 seconds.

If scan interval of "0:00:00" is specified, continuous scanning will occur when the Scan function is activated. If more than a few seconds are required to scan all defined channels, a short scan interval (a few seconds) effectively becomes continuous scanning.

If the Monitor function is turned on when the instrument is continuously scanning, a Monitor channel measurement is still taken between each scan.

Table 3-10 illustrates the button sequence used for setting scan interval. If necessary, refer to "Entering and Changing Numeric Values" for a more detailed description of the number editing technique used here.

 

 

Hour

 

 

 

 

 

 

 

 

 

 

 

PRESS

 

I

 

 

 

J

THESE

 

 

 

G

BUTTONS:

 

 

 

D

TO

 

 

0:00:00

 

 

 

 

 

 

CHANGE

 

 

 

 

 

 

 

 

 

 

THE

 

 

 

 

 

 

INDICATED

 

 

 

 

 

 

DIGIT:

 

 

 

 

 

 

 

 

 

 

 

 

 

Table 3-10. Scan Interval

Minute

J J

G

G

D

D

0:00:00

0:00:0

 

0

Second

JE

G

G

D

D

0:00:0

0:00:0

0

0

Note: Go backward (from SECOND to MINUTE to HOUR) by pressing H.

Press Eat any time to accept the displayed scan interval and exit scan interval selection. Press C at any time to exit interval selection without storing any changes.

Selecting the Measurement Rate

The slow measurement rate provides the highest accuracy and resolution. A fast rate can be selected, but keep in mind that fast rate provides one less digit of resolution than does slow rate. Measurement rate selection is illustrated in Table 3-11.