Fluke 2620A, 2625A Operating the Instrument from the Front Panel, Measurement Connections

Models: 2625A 2620A

1 197
Download 197 pages 4.31 Kb
Page 63
Image 63

Operating the Instrument from the Front Panel

3

Measurement Connections

2-WIRE (2T) CONNECTION

SOURCE

(4-WIRE)

11 12 13 14 15 16 17 18 19 20

H L H L H L H L H L H L H L H L H L H L

SENSE

(4-WIRE)

H L

1

H L H L H L H L H L H L H L H L H L

2

3

4

5

6

7

8

9

10

 

 

 

 

 

 

 

 

RESISTANCE

 

 

 

 

 

 

 

 

OR

 

 

 

 

 

 

 

 

RTD SOURCE

USE H AND L TERMINALS FOR ANY CHANNEL.

CHANNEL 0 ON FRONT PANEL

CHANNELS 1 THROUGH 20 ON REAR

PANEL INPUT MODULE (CHANNEL 8 SHOWN HERE).

4-WIRE (4T) CONNECTION

SOURCE

(4-WIRE)

SENSE

(4-WIRE)

11 12 13 14 15 16 17 18 19 20

 

 

 

 

 

 

 

 

 

 

 

H L

 

H L

H L

H L

H L

H L

H L

H L

H L

H L

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

H L

 

H L

H L

H L

H L

H L

H L

H L

H L

H L

 

 

 

 

 

 

 

 

 

 

 

1 2 3 4 5 6 7 8 9 10

RESISTANCE

OR

RTD SOURCE

USE H AND L TERMINALS FOR TWO CHANNELS ON REAR PANEL INPUT MODULE. CONNECTIONS FOR CHANNEL 8 SHOWN HERE WITH CHANNEL 18 PROVIDING ADDITIONAL TWO CONNECTIONS.

FOR EACH 4-WIRE CONNECTION, ONE SENSE CHANNEL (1 THROUGH 10) AND

ONE SOURCE CHANNEL (SENSE CHANNEL NUMBER +10 = 11 THROUGH 20) ARE USED.

oo09f.eps

Figure 3-3. 2-Terminal and 4-Terminal Connections

3-21