Fluke 2620A, 2625A Trigger Input, Appendices, Isolation, Threshold, Hysteresis, Input Voltages

Models: 2625A 2620A

1 197
Download 197 pages 4.31 Kb
Page 165
Image 165

Appendices

Specifications A

Isolation

none dc-coupled

Threshold

1.4V

Hysteresis

500 mV

Trigger Input

Input Voltages

contact closure and TTL compatible “high” =2.0V min, 7.0V max

“low” = -0.6V min, 0.8V max

Isolation

None dc-coupled

Minimum Pulse Width

5 µs

Maximum Frequency

5 Hz

Specified Conditions

The instrument must be in the quiescent state, with no interval scans in process, no commands in the queue, no RS-232 or IEEE interface activity, and no front panel activity if the latency and repeatability performance is to be achieved. For addition information, refer to Chapter 5.

Maximum Latency

Latency is measured from the edge of the trigger input to the start of the first channel measurement for the Specified Conditions (above).

480 ms for fast rate, scanning DCV, ACV, ohms, and frequency only

550 ms for fast rate, scanning any thermocouple or 100 mV dc channels 440 ms for slow rate, scanning DCV, ACV, ohms, and frequency only 890 ms for slow rate, scanning any thermocouple or 100 mV dc channels

Repeatability

3 ms for the Specified Conditions (above)

A-13