Fluke 2620A/2625A, Input Impedance, Maximum Crest Factor, Crest Factor Error, Maximum AC Input

Models: 2625A 2620A

1 197
Download 197 pages 4.31 Kb
Page 160
Image 160

2620A/2625A

Users Manual

Frequency

20 Hz - 50 Hz

50 Hz - 100 Hz

100 Hz - 10 kHz

10 kHz - 20 kHz

20 kHz - 50 kHz

50 kHz - 100 kHz

Maximum Input at Upper Frequency

300 V rms

300 V rms

200 V rms

100 V rms

40 V rms

20 V rms

Input Impedance

1 MΩ in parallel with 100 pF maximum Maximum

Maximum Crest Factor

3.0

2.0 for rated accuracy

Crest Factor Error

Non-sinusoidal input signals with crest factors between 2 and 3 and pulse widths 100 µs and longer add 0.2% to the accuracy specifications.

Common Mode Rejection

80 dB minimum at 50 or 60 Hz ± 0.1%, 1 kΩ imbalance, slow rate

Maximum AC Input

300V rms or 424V peak on channels 0, 1, and 11

150V rms or 212V peak on channels 2 to 10 and 12 to 20 Voltage ratings between channels must not be exceeded

2 x 106 Volt-Hertz product on any range, normal mode input

1 x 10 6 Volt-Hertz product on any range, common mode input

DC Component Error

SCAN and first MONitor measurements will be incorrect if the dc signal component exceeds 60 counts in slow rate or 10 counts in fast rate. To measure ac with a dc component present, MONitor the input and wait 5 seconds before recording the measurement.

Cross-Talk Rejection

Refer to Appendix D.

A-8

Page 160
Image 160
Fluke 2620A/2625A, Input Impedance, Maximum Crest Factor, Crest Factor Error, Common Mode Rejection, Maximum AC Input