Fluke 2620A, 2625A Using the Computer Interface, General Information RS-232 and IEEE-488

Models: 2625A 2620A

1 197
Download 197 pages 4.31 Kb
Page 85
Image 85

Using the Computer Interface

4

General Information (RS-232 and IEEE-488)

FUNC 1, OHMS, 3, 2

Select ohms function for channel 1

Select 30-kilohm range

Select 2-wire (2T) connections

FUNC 12, TEMP, K

Select temperature measurement for channel 12

Select K-type thermocouple input

FUNC 1, TEMP, PT, 2; RTD_R0 12, 101.22

Select temperature measurement for channel 1

Select RTD (DIN/IEC 751)

Select 2-wire (2T) connections

Set new R0 value for channel 12

INTVL 0, 10, 0; SCAN 1; LAST?

Set interval between scan starts to 10 minutes

Start scanning

Return measured values for all scanned channels

oo13f.eps

Figure 4-3. Typical Input Strings

4-15

Page 85
Image 85
Fluke 2620A Using the Computer Interface, General Information RS-232 and IEEE-488, FUNC 1, OHMS, 3, FUNC 12, TEMP, K