2620A, 2625A

Users Manual

through both a known resistance and the sensed resistance. The resulting voltages are measured and appropriate conversions are applied to the measurement, yielding a displayed output in ohms.

Frequency is measured by counting cycles for a known time period. The measurement represents the frequency observed during the sampling time. The instrument can measure a wide range of frequency inputs. Test applications might include measuring line voltage sine wave signals or measuring the output of a voltage-to-frequency converter used in a servo system.

Thermocouple temperature measurements can be made using linearizations for the following nine standard thermocouples: J, K, E, T, N, R, S, B, C. You specify what type of thermocouple is connected to the channel. The reference temperature sensor is built into the Input Module. The instrument applies compensation automatically for thermocouple channels. Open thermocouple detection is indicated by "otc" in the left display. The thermocouples are further described in Table 3-14.

 

 

 

 

Table 3-2. DC Voltage, AC Voltage

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Function

 

 

 

Range

 

 

 

 

ChannelG

 

 

 

(DC orGAC)

 

 

 

(Note)G

 

 

PRESS

 

 

 

F

 

 

 

E

 

 

 

E

THESE

 

 

 

 

 

 

 

 

 

BUTTONS:

 

D

 

 

 

D

 

 

 

D

 

 

 

 

 

 

 

 

 

 

 

 

 

 

TO

0

 

 

 

OFF

 

 

 

Auto

 

(Completes

SELECT

1

 

 

 

V DC

 

 

 

300.00 mV

 

Selection

FROM

.

 

 

 

V AC

 

 

 

3.0000V

 

and returns

THESE

.

 

 

 

Ω

 

 

 

30.000V

 

to Inactive

CHOICES:

 

 

 

 

 

Hz

 

 

 

150.00*V

 

Mode)

 

20

 

 

 

°C or °F

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

* 300.00 CH 0, 1, and 11

Note: Determine the highest ac or dc voltage value anticipated for this channel. Then select a range large enough to accommodate this value. If the highest voltage cannot be anticipated, select “Auto”.

3-6