Fluke user manual 2620A, 2625A, 8. Command and Query Reference cont

Models: 2625A 2620A

1 197
Download 197 pages 4.31 Kb
Page 102
Image 102

2620A, 2625A

Users Manual

Table 4-8. Command and Query Reference (cont)

FORMAT?

FUNC

MEASUREMENT

UNITS STRING

Scaled “MX+B”

 

Volts DC

“VDC”

Volts AC

“VAC”

Resistance

“OHMS”

Frequency

“Hz”

Temperature

“C”

Temperature

“F”

Query Response Format

Returns the format presently in use:

The default. IEEE-488.2-compatible. No measurement units are allowed.

Responses are RS-232-compatible, allowing a means of recording the units with the measurement value.

Channel Function Definition

Define the measurement function and range for the indicated channel. Attempting to use the FUNC command when measurements are active results in an Execution Error.

Note

Successful execution of the FUNC command clears any alarm limits and scaling values for this channel. Therefore, you must define a channel’s function before setting alarm limits and/or scaling values for that channel. Also, the FUNC command clears all values in the Review array. FUNC <channel>,<OFF>

FUNC <channel>,<VAC VDC FREQ>,<range>

FUNC <channel>,<OHMS>,<range>,<terminals>

FUNC <channel>,<TEMP>,<thermocouple type>

FUNC <channel>,<TEMP>,<PT>,<terminals>

<channel> = (0 .. 20)

If the channel number is invalid, an Execution Error is generated.

<function> = VAC VDC OHMS FREQ TEMP OFF

If the function is OFF, any additional supplied parameters generate a Command Error.

<range> = AUTO (1 .. 6) (Thermocouple/RTD_type)

Ranges for VAC, VDC, OHMS, and FREQ are listed in the following table. Enter AUTO to select autoranging. Use of any other values causes an Execution Error.

RANGE

VOLTAGE

OHMS

FREQUENCY

1

300 mV

300Ω

900 Hz

2

3 V

3kΩ

9 kHz

3

30 V

30 kΩ

90 kHz

4

150/300 V

300 kΩ

900 kHz

5

Error

3 MΩ

1 Mhz

6

Error

10 MΩ

Error

4-32

Page 102
Image 102
Fluke user manual 2620A, 2625A, 8. Command and Query Reference cont