Fluke user manual 2620A, 2625A, 8. Command and Query Reference cont

Models: 2625A 2620A

1 197
Download 197 pages 4.31 Kb
Page 104
Image 104

2620A, 2625A

Users Manual

Table 4-8. Command and Query Reference (cont)

IEE?

IER?

INTVL

INTVL?

LAST?

Instrument Event Enable Query

Returns the present value of the Instrument Event Enable Register as an integer.

Instrument Event Register Query

Returns the value of the Instrument Event Register as an integer, then clears all bits.

Set Scan Interval

Set scan interval time.

INTVL <hours>,<minutes>,<seconds>

<hours> = (0 .. 9) <minutes> = (0 .. 99) <seconds> = (0 .. 99)

An Execution Error is generated if values outside the specified ranges are used or if measurements are active.

Scan Interval Query

Return scan interval time. Returns three values: hours, minutes, and seconds.

Channel’s Last Scan Value

Returns value(s) for channels measured in the most recent scan. The value returned represents data from the most recent scan. The most recent scan is the scan in progress or, if scanning is not in progress, the last completed scan.

LAST? <channel> <channel> = (0 .. 20)

Returns measurement values for either the indicated channel or all defined channels. If the channel specification field is left blank, values for all defined channels are returned (values for channels defined as OFF are not included.) An Execution Error results if a request is made for a channel defined as OFF, the channel specified is invalid, the channel specified has been set up but not yet measured, or Review array values have been cleared.

The response is a signed number with decimal point and exponent. For slow scanning rate, 5 digits are returned (+/-XX.XXXE+/-X); for fast scanning rate, 4 digits are returned (+/-XX.XXE+/-X). The range setting determines placement of the decimal point.

For values of all defined channels (no channel specification made), undefined (OFF) channels are not included. For each defined channel, a separate signed number with decimal point and exponent is returned. Commas separate numbers for different channels, and no blank spaces are included.