Fluke 2620A Using the Computer Interface, Computer Interface Command Set, limitnum =, DOstate

Models: 2625A 2620A

1 197
Download 197 pages 4.31 Kb
Page 99
Image 99

Using the Computer Interface

4

Computer Interface Command Set

Table 4-8. Command and Query Reference (cont)

ALARM_ASSOC?

Alarm Association Query

 

Returns alarm output associations for the indicated channel and alarm

 

limit.

 

 

 

ALARM_ASSOC? <channel>,<limit_num>

 

<channel>

=

(4 .. 20)

 

<limit_num> =

1 2

 

Returns the digital output line number associated with the indicated

 

alarm limit. An Execution Error is generated if no output is associated

 

with this alarm limit, if invalid channel numbers are used, or if limits other

 

than 1 or 2 are used.

ALARM_ASSOC_CLR

Clear Alarm Association

 

Clear the digital output association for the indicated channel and alarm

 

limit.

 

 

 

ALARM_ASSOC_CLR <channel>,<limit_num>

 

<channel>

 

= (4 .. 20)

 

<limit_num> = 1 2

 

When the association is cleared, the output pin is left in a high (non-

 

alarm) state. If the alarm limit specified is not 1 or 2, or if the channel is

 

invalid, an Execution Error is generated.

 

 

ALARM_DO_LEVEL

Set Alarm Output Level

 

Set or clear the indicated alarm digital output.

 

ALARM_DO_LEVEL <DO_line>,<DO_state>

 

<DO_line> = (0 .. 3)

 

<DO_state>

= 1 (high) 0 (low)

 

Alarm outputs 0-3 correspond to channels 0-3, respectively. If the digital

 

output line requested is not in the range 0 through 3, an Execution Error

 

is generated. If the DO_state specified is not 1 (high) or 0 (low), an

 

Execution Error is generated.

 

 

ALARM_DO_LEVELS?

Alarm Output State Query

 

Returns digital output levels for the four alarm digital outputs.

Returns an integer value representing the state of each of the digital I/O lines. The low order four bits are the status of the alarm digital outputs; 0 indicates line is low (in alarm), and 1 indicates line is high (not in alarm). These lowest four bits correspond to alarm digital outputs 0 through 3, which are permanently associated to channels 0 through 3, respectively. A value of 15 indicates that all alarm outputs are in the non-alarm (high) state.

4-29

Page 99
Image 99
Fluke 2620A, 2625A user manual Using the Computer Interface, Computer Interface Command Set, limitnum =, DOstate