Fluke user manual Frequency Inputs, 2620A/2625A, Frequency Range, Sensitivity, Maximum AC Input

Models: 2625A 2620A

1 197
Download 197 pages 4.31 Kb
Page 162
Image 162

2620A/2625A

Users Manual

Frequency Inputs

Frequency Range

15 Hz to greater than 1 Mhz

Range

Resolution

 

 

Accuracy + (% ± Hz)

 

Slow

Fast

 

 

 

15 Hz - 900 Hz

0.01 Hz

0.1 Hz

 

 

 

9 kHz

0.1 Hz

1 Hz

 

 

 

90 kHz

1 Hz

10 Hz

 

 

 

900 kHz

10 Hz

100 Hz

 

 

 

1 MHz

100 Hz

1 kHz

 

 

 

Sensitivity

Slow

0.05% + 0.02 Hz

0.05% + 0.1 Hz

0.05% + 1 Hz

0.05% + 10 Hz

0.05% + 100 Hz

Fast

0.05% + 0.2 Hz

0.05% + 1 Hz

0.05% + 10 Hz

0.05% + 100 Hz

0.05% + 1 kHz

Frequency

15 Hz - 100 kHz

100 kHz - 300 kHz

300 kHz - 1 MHz Above 1 MHz

Level (Sine Wave)

100 mV rms

150 mV rms

2 V rms

Not specified

Maximum AC Input

300V rms or 424V peak on channels 0, 1, and 11

150V rms or 212V peak on channels 2 to 10 and 12 to 20 Voltage ratings between channels must not be exceeded

2 x 106 Volt-Hertz product on any range, normal mode input

1 x 106 Volt-Hertz product on any range, common mode input

Cross-Talk Rejection

Refer to Appendix D.