Fluke 2625A, 2620A user manual RTD Temperature Accuracy Test

Models: 2625A 2620A

1 197
Download 197 pages 4.31 Kb
Page 142
Image 142

2620A, 2625A

Users Manual

RTD Temperature Accuracy Test

The following two RTD Temperature Accuracy Tests are different in that one uses a Decade Resistance Source and the other uses an RTD. Only one of the tests need to be performed to assure operation.

RTD Temperature Accuracy Test (Using Decade Resistance Source)

Assure Channel 0’s Accuracy Verification Test for DC Volts and 300 Ohm Resistance Range meets minimum acceptable levels.

1.Switch OFF power to the instrument and disconnect all high voltage inputs.

2.Remove the Input Module from the rear of the instrument. Open the Input Module and connect a pair of test leads (keep as short as possible) to the H (high) and L (low) terminals of channel 1. For 4-terminal performance testing, connect a second pair of test leads to the H (high) and L (low) terminals of channel 11. Install the Input Module back into the instrument.

3.Connect channel 1’s test leads to the Output HI and LO terminals of the Decade Resistance Source. For 4-terminal performance testing, also connect channel 11’s test leads to the Output HI and LO terminals of the Decade Resistance Source. Connect as shown in Figure 6-3.

Note

4-terminal connections are made using pairs of channels. 4-terminal measurements can only be made on channels 1 through 10. The accompanying pairs are channels 11 through 20.

4.Switch the instrument ON.

5.Select the 4-terminal RTD temperature function, RTD type PT, for channel 1 on the Hydra Series II Instrument. Press MON and ensure the display reads between the minimum and maximum values shown on Table 6-5.

6.The RTD Temperature Accuracy test is complete. However if you desire to perform this test on Input Module channels (2 through 10) repeat steps 1 through 5 substituting in the appropriate channel number.

Note

The only type of temperature measurement that can be made on channel 0 is 2-terminal RTD. Channels 11 through 20 will support only 2-terminal RTD’s.

Table 6-5. Performance Tests for RTD Temperature Function (Resistance)

(DIN/IEC 751 Amendment 1) (IPTS-68)

DECADE RESISTANCE

SOURCE

100e

200e

300e

TEMPERATURE SIMULATED

°C

0

266.42

558.00

TEMPERATURE ACCURACY

1YEAR @ 18-28°C

±0.12°C

±0.22°C

±0.37°C

These figures assume that RTD R0 is set to 100.00 ohms for each channel.

Accuracy given is for 4-wire measurements only. For 2-wire measurements, degrade the accuracy specifications by 5.2 °C per ohm of single lead wire resistance. For 2-wire measurements, degrade the accuracy by an additional 11°C (channels 1 .. 20) or 0.05°C (channel 0).

6-14

Page 142
Image 142
Fluke 2625A, 2620A user manual RTD Temperature Accuracy Test Using Decade Resistance Source